Comparison of electroretinographic measurements between tabletop and handheld stimulators in healthy subjects
Bum G Kim1, In B Chang2, Kyeong D Jeong3
1Department of Ophthalmology, Sanggye Paik Hospital, Inje University College of Medicine, Seoul, Korea.
Purpose:
To compare full-field electroretinography (ffERG) parameters obtained from handheld and tabletop electroretinography (ERG) devices in normal subjects.
Methods:
Twenty volunteers underwent ffERG using a tabletop and handheld stimulator. The responses obtained from the right eyes were compared. The coefficient of variation and intraclass correlation coefficient (ICC) were derived to assess inter- and intra-individual reliability.
Results:
The b-wave in the rod response, a- and b-waves in the maximal combined response, a-wave in the cone response, and the 30-Hz flicker response showed significantly greater amplitudes when recorded with the tabletop stimulator than with the handheld stimulator. The implicit time of response (ITR) in the 30-Hz flicker response was longer when recorded with the handheld stimulator than when recorded with the tabletop stimulator. With regard to amplitude, the ICC indicated moderate-to-high reliability in the measurement of the b-wave in the rod response, and a- and b-waves in the maximal combined response. With regard to ITR, measurement of the b-wave in the rod response and a-wave in the maximal combined response showed moderate-to-high reliability.
Conclusion:
Despite the significantly lower ERG amplitude measurements recorded by the handheld stimulator, there were no significant differences in variability between the two stimulators.
Related Concept Videos
Bioavailability Study Design: Healthy Subjects Versus Patients
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Cable Subjected to Its Own Weight
A generalized loading function is employed to analyze a cable subjected to its own weight. This function considers the force acting along the cable's arc length rather than its projected length, providing a more accurate...
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Assessment of the Gastrointestinal System I: Subjective Data
Health History
The initial step in assessing the GI system is obtaining a comprehensive health history. This includes inquiring about the patient's history or presence of problems...
Cable Subjected to Concentrated Loads