Image quality and diagnostic value of ultra low-voltage, ultra low-contrast coronary CT angiography
Chong-Fu Jia1, Jie Zhong2, Xin-Yi Meng3
1Department of Cardiovascular Radiology, The First Affiliated Hospital of Dalian Medical University, 222#, Zhong Shan Road, Dalian, 116011, Liaoning, China.
Objective:
To explore the image quality (IQ) and diagnostic value of 70 kVp turbo high-pitch coronary CT angiography (THP-CCTA) using automated tube voltage selection (ATVS) and 30 mL of low-concentration contrast agent.
Methods:
Patients who underwent 70 kVp THP-CCTA using ATVS with 30 mL of contrast agent (group A) were prospectively enrolled, and those who underwent conventional CCTA (100/120 kVp, prospective sequential mode with 65-75 mL of contrast agent) (group B) were retrospectively selected for study. IQ was assessed subjectively on a 5-point scale, and diagnostic value was assessed based on invasive coronary angiography as the gold standard. Heart rate (HR), HR fluctuation (HRF), body mass index (BMI), effective radiation dose (ED), and iodine uptake (IU) were recorded.
Results:
A total of 796 patients (398/398 in groups A/B) were included. Between-group differences in age, gender, BMI, HR, HRF, and IQ values were not significant. The ED/IU values were 0.3 ± 0.1 mSv/9.0 ± 0.0 g and 5.8 ± 1.8 mSv/22.9 ± 1.0 g in groups A and B, respectively (p < 0.01). The sensitivity, specificity, positive and negative predictive values, and accuracy of THP-CCTA for the diagnosis of ≥ 50% stenosis were 94.8%, 97.5%, 92.0%, 98.4%, and 96.9% respectively. The mean HR and coronary calcium score were independent predictors of diagnostic image quality, and the best cutoff values were 71.5 bpm and 444.1 respectively.
Conclusion:
This third-generation dual-source CT imaging modality, a 70-kVp THP-CCTA system using ATVS with 30 mL of low-concentration contrast agent, produces high-quality images with high diagnostic accuracy for significant stenosis, with ultra low ED and IU. This technique was most promising in individuals with an HR < 71.5 bpm and coronary calcium score < 444.1.
Key Points:
• Turbo high-pitch CCTA using 70 kVp via automated tube voltage selection and 30 mL of low-concentration contrast agent is feasible. • This protocol provides high diagnostic accuracy for significant coronary stenosis and reduces radiation doses and iodine uptake significantly. • This protocol was most promising in individuals with an HR < 71.5 bpm and coronary calcium score < 444.1.
Related Concept Videos
Acute Coronary Syndrome III: Diagnostic Studies
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Voltage
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...