Variations in the Efficiency of Albuterol Delivery and Intrapulmonary Effects With Differential Parameter Settings on
Takashi Karashima1,2, Yuka Mimura-Kimura2, Kanade Miyakawa1
1Department of Medical Engineering, National Hospital Organization Yamaguchi-Ube Medical Center, Ube, Japan.
Background:
Intrapulmonary percussive ventilation (IPV) is used for airway clearance and delivery of aerosol medications, including bronchodilators. Despite the common use of IPV for drug delivery, few data are available regarding optimization of inhalation therapy with IPV. In this study, we investigated the influence of IPV setting parameters and lung mechanics on drug delivery via IPV alone.
Methods:
An IPV device was connected to a lung model via a trachea model and a flow analyzer. Albuterol nebulized from the IPV device was collected onto a filter attached between the trachea and lung models, and was quantitated by spectrophotometry (230 nm).
Results:
Albuterol delivery to the lung model was increased up to 2.1-fold, with decreasing percussion frequency. Decreasing percussion frequency concomitantly increased the tidal volume, and albuterol delivery was correlated with tidal volume (r = 0.91, P < .001). Airway resistance had a negative impact on albuterol delivery, whereas lung compliance had no significant effect. Increasing operational pressure increased albuterol delivery while increasing peak inspiratory pressure.
Conclusions:
Albuterol delivery and tidal volume with IPV can be improved by maintaining low levels of percussion frequency and increasing operational pressure. When increasing operational pressure, the peak inspiratory pressure and airway resistance levels need to be carefully monitored for safe inhalation therapy with IPV.
Related Concept Videos
Mechanical Ventilation I: Indication and Settings
What is Variation?
The range, standard deviation, standard error, and variance are the different measures of variation.
Range: The range is the difference between its maximum and...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Variation
When independent and dependent variables are plotted on a scatter plot, the slope of a line is a value that describes the rate of change between the two...
Assessment of the Abdomen II: Percussion
Percussion
Percussion is an essential...