[Evaluation of Low-contrast Detectability of Iterative Reconstruction Algorithm in High-speed CT Technology for Head]
Shota Masuda1, Koichi Sugisawa1, Kazuya Minamishima1
1Department of Radiological Technology, Keio University Hospital.
Purpose:
The purpose of this study was to assess the feasibility of high-speed CT technology for head without deterioration of low-contrast detectability using the brain LCD (Canon Medical Systems) of iterative reconstruction.
Methods:
System performance (SP) function analysis, low-contrast object specific contrast-to-noise ratio (CNRLO) analysis, and visual evaluation using Scheffe's paired comparison were performed. Additionally, analysis of the correlation of CNRLO and visual scores was performed. SP was performed with the self-made phantom. CNRLO was calculated with the catphan 504 phantom (CTP 515). Visual evaluation was performed using the brain phantom which simulated such as cerebral infarction and investigated on a fivepoint scale. All images were acquired with pitch factor of 0.61 (low pitch) and 1.40 (high pitch). All images were reconstructed with filtered back projection (FBP), brain LCD standard (LCD STD) and strong (LCD STR).
Results:
SP of brain LCD improved compared with FBP. CNRLO of FBP decreased in high pitch compared with low pitch. CNRLO of brain LCD images acquired by low- and high pitch were improved compared with FBP. Visual scores denoted similar trends to that of CNRLO and there was high correlation with CNRLO.
Conclusion:
It was suggested that using brain LCD can achieve the high speed CT technology for head without deterioration of low-contrast detectability.
Related Concept Videos
Trial and Error and Algorithm
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Muscles that Move the Head
The bilateral sternocleidomastoid, or SCM, and the suprahyoid and infrahyoid muscles are significant head flexors. The SCM muscles originate at the sternum and clavicle and attach to the mastoid process of the temporal bone. The SCM contracts bilaterally to bend the head forward, whereas...
Speed of a Transverse Wave
One of the key properties of any wave is the wave speed. Light...
Speed of Sound in Gases
Distribution of Molecular Speeds