Primed Left Ventricle Heart Perfusion Creates Physiological Aortic Pressure in Porcine Hearts
Liming Xin1,2, Weiran Yao1,3, Yan Peng4
1From the Department of Mechanical & Industrial Engineering, University of Toronto, Toronto, Ontario Canada.
Abstract:
This article presents a primed left ventricle heart perfusion method to generate physiologic aortic pressure (AoP) and perform functional assessment. Isolated hearts of male Yorkshire pigs were used to study the hemodynamic behaviors of AoPs generated in the primed left ventricle heart perfusion (n = 6) and conventional (zero-loaded left ventricle) Langendorff perfusion (n = 6). The measurement results show that left ventricular pressure generated in the primed left ventricle heart perfusion is a determinant of physiologic AoP (i.e. systolic and diastolic pressures within physiologic range). The aortic pulse pressure (systolic pressure = 124.5 ± 1.7 mm Hg, diastolic pressure = 87.8 ± 0.9 mm Hg, aortic pulse pressure = 36.7 ± 2.6 mm Hg) from the primed left ventricle heart perfusion represents close match with the in vivo physiologic data. The volume in the left ventricle remains constant throughout the primed left ventricle heart perfusion, which allows us to perform isovolumetric left ventricular pressure measurement in ex vivo heart perfusion (EVHP). Left ventricular contractility measurements (maximum and minimum rates of left ventricular pressure change) were derived for cardiac assessment. In summary, the proposed primed left ventricle heart perfusion method is able to create physiologic AoP and enables left ventricular functional assessment in EVHP in porcine hearts.
Related Concept Videos
Anatomy of the Heart
Anatomy of the Heart
The heart has three layers: the innermost endocardium, the muscular myocardium, and the outer epicardium, all working together for optimal cardiac function.
Chambers of the Heart
The heart is made up of four...
Physiology of the Heart: The Cardiac Cycle
Diastole: The Relaxation Phase
During diastole, all four heart chambers relax. The atrioventricular (AV) valves open, and the semilunar valves close. This phase sees the lowest chamber pressures, promoting ventricular filling. Venous blood enters the heart through the...
Cardiac Catheterization III: Left Heart Catheterization
Overview of the Heart
The heart's structure...
Conduction System of the Heart
The pacemaker cells are located in two primary nodes: the sinoatrial (SA) node and the atrioventricular (AV) node. The SA node pacemaker cells can autonomously depolarize, triggering an action potential that leads to the...