Portable fixed dynamometry to quantify ankle dorsiflexion force
Marita E Mauracher1, Michael J Asmussen2, Sandro R Nigg2
1Institute for Biomechanics, ETH Zurich, Zurich, Switzerland.
Introduction:
Quantifying muscle strength is critical in clinical and research settings. A rapid and objective method is ideal. The primary objective of this study was to examine the reliability of a novel device, the rapid objective quantification- tibialis anterior (ROQ-TA), which quantifies the dorsiflexion force of the tibialis anterior, and to assess its validity against isokinetic dynamometry (IKD).
Methods:
Ankle dorsiflexion of 20 healthy subjects was assessed by 3 modalities, ROQ-TA, manual muscle testing, and isokinetic dynamometry, over 2 testing sessions.
Results:
The intraclass correlation coefficient [ICC(2,1) ] for reliability was 0.872 (0.677-0.949) for the ROQ-TA and 0.892 (0.728-0.957) for IKD. For validity, the ICC(2,1) values for the ROQ-TA and IKD were in good agreement, with 0.672 (0.17-0.87) in the first testing session and 0.769 (0.42-0.91) in the second session.
Discussion:
The ROQ-TA is a valid and reliable device to test ankle dorsiflexion force in a healthy population. Muscle Nerve, 2018.