External continuous ECG versus loop recording for atrial fibrillation detection in patients who had a stroke
Michala Herskind Sejr1, Ole May1, Dorte Damgaard2
1Department of Cardiology, Cardiovasculair Research Unit, Regional Hospital West Jutland, Herning, Denmark.
Background:
Detection of atrial fibrillation (AF) in patients who had ischaemic stroke and transient ischaemic attack (IS/TIA) is recommended. We aimed to compare external loop recording (ELR) against simultaneous continuous ECG recording for AF detection in patients who had acute IS/TIA and determine sensitivity, specificity and positive predictive value of AF detection using ELR. We hypothesised ELR to detect 15% fewer patients with AF than continuous ECG recording.
Methods:
In this prospective cohort study, we included 1412 patients who had acute IS/TIA without prior AF. Monitoring was 48 hours. Primary outcome was AF >30 s. Cardiologist verified AF in continuous ECG was gold standard.
Results:
In continuous ECG, 38 (2.7%) patients had AF. ELR automatically categorised 219/1412 patients (15.5%) with AF, including 32/38 (85%) patients with AF in continuous ECG. After cardiologist adjudication of ELR recordings, AF was diagnosed in 57/219 patients, of which 32 (56%) had AF in continuous ECG. For adjudicated AF detection by ELR, sensitivity was 84%, 95% CI (69% to 94%), specificity was 98%, 95% CI (97% to 99%) and positive predictive value was 56%, 95% CI (42% to 69%).
Conclusion:
Automatic AF detection with ELR results in an AF diagnosis in more than five patients without AF for each patient with AF as verified in continuous ECG. For adjudicated AF detection by ELR, sensitivity was confirmed to 84% and specificity 98%. Automatic ELR as investigated in this study may be considered to rule out AF, but it is not suitable as a single monitoring device for AF screening in patients early after stroke.
Trial Registration Number:
NCT02155907.
Related Concept Videos
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Amyloid Fibrils
Amyloid deposits were observed as early as 1639 in the liver and the spleen. In 1854, Rudolph Virchow performed iodine staining,...
Amyloid Fibrils
ECG Interpretation of Rhythms
Components of the Electrocardiogram
The primary components of a normal ECG waveform in Normal sinus rhythm(NSR) include the P wave, PR interval, QRS complex, ST segment, T wave, and occasionally a U wave.
ECG waveforms are divided by vertical and horizontal lines at standard intervals.
The horizontal axis measures time and rate, and the vertical axis measures amplitude or voltage....
Fibril-associated Collagen
For example, the type II collagen fibrils in cartilage have covalently bound type IX fibril-associated collagens at regular intervals. Other types of fibril-associated collagens are...
Work Done on a System by External Force
In the presence of a non-conservative opposing force, like friction, some part of the work done...