Comparison of central corneal thickness measurements using different imaging devices and ultrasound pachymetry
Ertugrul Can1, Hilal Eser-Ozturk1, Mustafa Duran2
1Department of Ophthalmology, Ondokuz Mayis University Faculty of Medicine, Samsun, Turkey.
Purpose:
To compare central corneal thickness (CCT) measurements obtained by the AL-Scan, Lenstar LS900, Galilei, and ultrasound pachymetry (UP) in normal and cataractous eyes.
Methods:
Eighty eyes of healthy subjects were included in the study. Each subject was assessed by four different methods of measurements using the AL-Scan, Lenstar LS900, Galilei, and UP by a single examiner. To assess the intraobserver repeatability, three consecutive measurements were taken for the AL-Scan.
Results:
The mean CCT [± standard deviation (SD)] for the AL-Scan, Lenstar LS900, Galilei, and UP were 554.6 ± 30.9 μm, 542.9 ± 31.3 μm, 570.7 ± 30 μm, and 552.7 ± 32.8 μm, respectively. The differences between pairs of mean CCT for the methods are statistically significant for the pairs of Galilei-UP, AL-Scan-Galilei, and Lenstar LS900-Galilei. Bland-Altman plots showed that AL-Scan-UP have the closest agreement, followed by Lenstar-UP and AL-Scan-Lenstar. Galilei was found to have the poorest agreement with the other three methods. The intraobserver repeatability of the AL-Scan was very good with an intraclass correlation coefficient (ICC) of 0.980.
Conclusion:
We found that CCT measurements between the AL-Scan-UP, Lenstar LS900-UP, and AL-Scan-Lenstar LS900 showed very strong correlation and comparable agreement. AL-Scan-UP showed the closest agreement and these devices can be used interchangeably in clinical practice. Galilei significantly showed higher value of CCT compared to other methods. It was also observed that the Al-Scan had excellent intraobserver repeatability.
More Related Videos
10:39Fabrication and Characterization of Thickness Mode Piezoelectric Devices for Atomization and Acoustofluidics
Published on: August 5, 2020
04:48Establishing a Severe Corneal Inflammation Model in Rats Based on Corneal Epithelium Curettage Combined with Corneal Sutures
Published on: November 22, 2024
Related Concept Videos
Measures of Central Tendency
Peripherally and Centrally Acting Muscle Relaxants: A Comparison
Centrally acting muscle relaxants can be further divided into spasmolytic and antispasmodic drugs. Spasmolytic...
The Central Dogma
The Central Dogma
RNA is the Missing Link Between DNA and Proteins
In the early 1900s, scientists discovered that DNA stores all the information needed for cellular functions and that proteins perform most of these functions. However, the mechanisms of converting genetic information into functional proteins remained unknown for many years. Initially, it was believed that a single gene is...
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...