Related Experiment Video
Updated: Jun 28, 2026

Utilizing Percutaneous Ventricular Assist Devices in Acute Myocardial Infarction Complicated by Cardiogenic Shock
Published on: June 12, 2021
105 Inappropriate ICD Shocks in a Patient with Dilated Cardiomyopathy and Broca's Aphasia
Christian Georgi1, Michael Neuß1, Viviane Möller1
1Heart Center Brandenburg-Department of Cardiology and Medical School Brandenburg Theodor Fontane, Bernau bei Berlin, Germany.
Abstract:
With a growing number of ICD recipients, device complications are seen more frequently in the clinical setting and outpatient departments. Among the most severe are ICD infections and inappropriate therapies caused by oversensing of atrial tachycardias or lead fracture. We report on a 76-year-old female patient with dilative cardiomyopathy and Broca's aphasia after stroke, who experienced 105 consecutive inappropriate ICD shocks due to cluster missensing of her fractured ICD lead. The diagnosis was complicated and delayed by patient's aphasia emphasizing the need for intensified remote monitoring along with regular in-person visits, especially in people with intellectual or communication disabilities.
Related Concept Videos
Heart Failure Drugs: Inotropic Agents
Acute Coronary Syndrome IV: Interprofessional Care
Cardiopulmonary Resuscitation IV: Pharmacological Management
Cardiomyopathy II: Dilated Cardiomyopathy
Cardiomyopathy III: Hypertrophic Cardiomyopathy
Cardiomyopathy V: Interprofessional Care