Radiosynthesis, Quality Control, and Small Animal Positron Emission Tomography Imaging of 68Ga-Labelled Nano Molecules
Published on: October 4, 2024
Multicenter integrated positron emission tomography/computer tomography imaging quality assurance accreditation
Pan-Fu Kao1,2,3, Kuo-Wei Lee4, Chi-Wen Huang2,3
1Institute of Medicine, Chung Shan Medical University, Taichung, Taiwan, ROC.
Background:
Integrated positron emission tomography/computer tomography (PET/CT) image quality assurance ensures accurate, reproducible, and quantitative assessment of comparable scanner performance. We performed a preliminary multicenter PET/CT imaging quality assurance test with a fillable tomographic phantom in six medical centers in Taiwan.
Methods:
The phantom was filled with fixed proportions of fluorine-18 radionuclide solution in the background and with different spheres to simulate cold and hot lesions, and body background radioactivity. Imaging acquisitions were performed by using recommended parameters in different sites according to different brand names of the instrument. All imaging was subjectively scored by eight experienced nuclear medicine physicians as the spatial resolution of four hot vials (score 0-4), six cold spheres (score 0-6), and six cold rod areas (score 0-6), and overall satisfaction (score 0-5). Interobserver correlation and receiver operating characteristic (ROC) curve were analyzed.
Results:
The detection ability of hot vials, cold spheres, and cold rods was 4.0 ± 0.1, 5.2 ± 0.8, and 3.8 ± 0.9, respectively. Overall satisfaction was 4.0 ± 0.8. The ROC analysis revealed that the area under the curve for hot vials, cold spheres, and cold rods was 0.984, 0.887, and 0.928 respectively. The interobserver correlation for detectability of cold spheres and cold rods was 0.88 and 0.96, respectively.
Conclusion:
The results of the study indicated that (1) PET/CT imaging quality assurance for comparable scanner performance could be established on the basis of a standard phantom and (2) good interobserver correlation can be observed for those with accurate and interpretable results.
More Related Videos
08:36Multi-Tracer Studies of Brain Oxygen and Glucose Metabolism Using a Time-of-Flight Positron Emission Tomography-Computed Tomography Scanner
Published on: June 7, 2024
10:20Automation of a Positron-emission Tomography PET Radiotracer Synthesis Protocol for Clinical Production
Published on: October 26, 2018
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Quality Assurance
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Reliability and Validity