The accuracy of computed tomography scans for rapid prototyping of canine skulls
Michaela L Comrie1,2, Gabrielle Monteith3, Alex Zur Linden3
1Department Human Health and Nutritional Science, College of Biological Science, University of Guelph, Guelph, Ontario, Canada.
Abstract:
This study's objective was to determine the accuracy of using current computed tomography (CT) scan and software techniques for rapid prototyping by quantifying the margin of error between CT models and laser scans of canine skull specimens. Twenty canine skulls of varying morphology were selected from an anatomy collection at a veterinary school. CT scans (bone and standard algorithms) were performed for each skull, and data segmented (testing two lower threshold settings of 226HU and -650HU) into 3-D CT models. Laser scans were then performed on each skull. The CT models were compared to the corresponding laser scan to determine the error generated from the different types of CT model parameters. This error was then compared between the different types of CT models to determine the most accurate parameters. The mean errors for the 226HU CT models, both bone and standard algorithms, were not significant from zero error (p = 0.1076 and p = 0.0580, respectively). The mean errors for both -650HU CT models were significant from zero error (p < 0.001). Significant differences were detected between CT models for 3 CT model comparisons: Bone (p < 0.0001); Standard (p < 0.0001); and -650HU (p < 0.0001). For 226HU CT models, a significant difference was not detected between CT models (p = 0.2268). Independent of the parameters tested, the 3-D models derived from CT imaging accurately represent the real skull dimensions, with CT models differing less than 0.42 mm from the real skull dimensions. The 226HU threshold was more accurate than the -650HU threshold. For the 226HU CT models, accuracy was not dependent on the CT algorithm. For the -650 CT models, bone was more accurate than standard algorithms. Knowing the inherent error of this procedure is important for use in 3-D printing for surgical planning and medical education.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Concepts and Prototypes
The brain organizes this information using concepts, which are mental categories grouping linguistic data,...
Improving Translational Accuracy
Uncertainty in Measurement: Accuracy and Precision
Sutures of the Skull
Sutures are immobile joints between adjacent bones of the skull. The narrow gap between the bones is filled with dense, fibrous connective tissue that unites the bones. The long sutures located between the skull bones are not straight but instead follow irregular, tightly twisting paths. These twisting lines tightly...