Fluorescence emission difference microscopy by superoscillation excitation

Yong Liu1,2, Xiaona Wang2, Shaocong Liu2

  • 1College of Electronics and Information Engineering, Shanghai University of Electric Power, Shanghai, China.

Journal of Microscopy
|March 26, 2019
PubMed

Related Concept Videos

Emission Spectra02:39

Emission Spectra

When solids, liquids, or condensed gases are heated sufficiently, they radiate some of the excess energy as light. Photons produced in this manner have a range of energies, and thereby produce a continuous spectrum in which an unbroken series of wavelengths is present.
75.9K
Atomic Spectroscopy: Absorption, Emission, and Fluorescence01:23

Atomic Spectroscopy: Absorption, Emission, and Fluorescence

Atomic spectroscopy is a vital tool in elemental analysis, both qualitatively and quantitatively. It can be broadly divided into optical spectroscopy, mass spectroscopy, and X-ray spectroscopy methods. The optical spectroscopic methods are atomic absorption spectroscopy (AAS), atomic emission spectroscopy (AES), and atomic fluorescence spectroscopy (AFS). The first step in all three methods is atomization, where the solid, liquid, or solution-phase samples are converted into gas-phase atoms and...
2.7K
Confocal Fluorescence Microscopy01:16

Confocal Fluorescence Microscopy

Confocal microscopy is an advanced microscopic technique. The prime advantage of the confocal microscope over other microscopy techniques is its ability to block the out-of-focus light from the illuminated samples using pinholes. It is widely used with fluorescence optics to obtain high-resolution, sharp contrast images. Unlike optical microscopes, confocal microscopes use a focused beam of light laser to scan the entire sample surface at different z-planes. These microscopes are, therefore,...
20.1K
Super-resolution Fluorescence Microscopy01:37

Super-resolution Fluorescence Microscopy

Super-resolution fluorescence microscopy (SRFM) provides a better resolution than conventional fluorescence microscopy by reducing the point spread function (PSF). PSF is the light intensity distribution from a point that causes it to appear blurred. Due to PSF, each fluorescing point appears bigger than its actual size, and it is the PSF interference of nearby fluorophores that causes the blurred image. Various approaches to achieving higher resolution through SRFM have recently been...
12.4K
Difference from Background: Limit of Detection01:05

Difference from Background: Limit of Detection

The limit of detection (LOD) is the smallest amount of analyte that can be distinguished from the background noise. The LOD value corresponds to the concentration at which the analyte signal is three times larger than the standard deviation of the blank signal. Below this value, the analyte signal cannot be differentiated from the background noise. It is calculated by dividing the calibration slope by 3 times the standard deviation of the blank signals.
The LOD indicates the presence or absence...
8.2K
Electric Potential and Potential Difference01:16

Electric Potential and Potential Difference

Suppose a positive test charge moves away from a positive static charge, then the Coulomb force does positive work, and its electric potential energy decreases. The potential energy per unit charge is defined as the electric potential. The electric potential is independent of the test charge.
When a test charge moves from the initial to the final position, the electric potential difference between those positions is defined as the ratio of the change in the potential energy to the charge on the...
5.6K