Algorithm-based reduction of inappropriate defibrillator shock: Results of the Inappropriate Shock Reduction wIth
Ricardo Ruiz-Granell1, Emilio Vincenzo Dovellini2, Antoine Dompnier3
1Hospital Clinico Universitario, Valencia, Spain.
Background:
Inappropriate shocks (IS) continue to have a major negative impact on patients implanted with defibrillators.
Objective:
The purpose of this study was to assess IS reduction with the PARAD+ discrimination algorithm in a general population implanted for primary or secondary prevention.
Methods:
ISIS-ICD (Inappropriate Shock Reduction wIth PARAD+ Rhythm DiScrimination-Implantable Cardioverter Defibrillator) was a 2-year international, interventional study in patients implanted with a dual implantable cardioverter-defibrillator (ICD) or triple-chamber defibrillator (cardiac resynchronization therapy-defibrillator [CRT-D]) featuring PARAD+. IS (shocks not delivered for ventricular tachycardia or fibrillation) were independently adjudicated. The primary endpoint was percentage of IS-free patients at 24 months. Primary and worst-case analyses of annual incidence rates of patients with ≥1 IS, overall and per defibrillator type, were conducted.
Results:
In total, 1013 patients (80.7% male; age 67.1 ± 11.4 years; 68%/30%/2% primary/secondary/other indication) were enrolled and followed for a median of 552 days (interquartile range 354; 725). Of 993 analyzed patients programmed with PARAD+, 14 had ≥1 IS, corresponding to a percentage free from IS of 98.1% (95% confidence interval [CI] 96.8%- 98.9%). Annual incidence rates (per 100 person-years) of patients with IS were 1.0 (95% CI 0.59-1.69) and 2.1 (95% CI 1.46-3.02) in the primary and worst-case analyses, respectively. In ICD patients, rates were 1.2 (95% CI 0.68-2.23) and 2.3 (95% CI 1.47-3.53), and in CRT-D patients 0.59 (95% CI 0.19-1.83) and 1.8 (95% CI 0.93-3.44) per 100 person-years.
Conclusion:
The annual rate of defibrillator patients with IS using the enhanced PARAD+ discrimination algorithm alone ranged from 1.0 to 2.1 per 100 person-years in a general population implanted for primary or secondary prevention.
More Related Videos
Related Concept Videos
Oxidation-Reduction Reactions
Shock Waves
When the source's speed approaches the speed of sound, constructive interference between successive wavefronts emitted by the source occurs immediately behind it. Initially, scientists believed that this constructive interference would result in such high...
Block Diagram Reduction
The first step in this process is the identification and relocation of a branch point. A branch point, where a...
Oxymercuration-Reduction of Alkenes
Blood Pressure Imbalances and Circulatory Shock
Blood Pressure: Hypertension and Hypotension
Normal blood pressure is 120/80 mm Hg. Elevated blood pressure is 120-129/under 80 mm Hg. Hypertension, warranting treatment at 130/80 mm Hg, is often asymptomatic and can lead to severe cardiovascular events, aneurysms, peripheral arterial disease, chronic renal disease, or cardiac...
Phase I Reactions: Reductive Reactions