Is CT bulletproof? On the use of CT for characterization of bullets in forensic radiology
L E Paulis1, J Kroll2, L Heijnens2
1Department of Radiology & Nuclear Medicine, Maastricht University Medical Centre, PO Box 5800, 6202AZ, Maastricht, The Netherlands. leonie.paulis@mumc.nl.
Purpose:
Forensic investigations could benefit from non-invasive in situ characterization of bullets. Therefore, the use of CT imaging was explored for the analysis of bullet geometry and composition. Bullet visualization with CT is challenging as the metal constituents suffer from excessive X-ray attenuation due to their high atomic number, density, and geometry.
Methods:
A metal reference phantom was developed containing small discs of various common metals (aluminum, iron, stainless steel, copper, brass, tungsten, and lead). CT images were acquired with 70-150 kVp and 200-400 mAs and were reconstructed using an extended Hounsfield unit (HU) scale (- 10,240 to + 30,710). For each material, the mean CT number (HU) was measured to construct a metal database. Different bullets (n = 11) were scanned in a soft tissue-mimicking phantom. Bullet size and shape were measured, and composition was evaluated by comparison with the metal database. Also, the effect of bullet orientation within the CT scanner was evaluated.
Results:
In the reference phantom, metals were classified into three groups according to their atomic number (Z): low (Z ≤ 13; HU < 3000), medium (Z = 25-30; HU = 13,000-20,000), and high (Z ≥ 74; HU > 30,000). External bullet contours could be accurately delineated. Internal interfaces between jacket and core could not be identified. Cross-sectional spatial profile plots of the CT number along bullets' short axis revealed beam hardening and photon starvation effects that depended on bullet size, shape, and orientation within the CT scanner. Therefore, the CT numbers of bullets were unreliable and could not be used for material characterization by comparison with the reference phantom.
Conclusion:
CT-based characterization of bullets was feasible in terms of size and shape but not composition.
Related Concept Videos
Radiological Investigation I: X-ray and CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...