Purpose:
This study aimed to determine setup accuracy using anatomic landmarks for breast irradiation with and without surface guided radiation therapy (SGRT) and assess setup time with SGRT.
Methods And Materials:
This study included 115 patients with 1945 treatment fractions. Patients were treated with 4 techniques: tangents, tangents using deep-inspiration breath hold, and tangents with regional nodal irradiation with and without deep-inspiration breath hold. A total of 915 portal verification images were analyzed to determine setup errors for the skin, chest wall (CW), and heart. Setup error at each landmark was defined as the mean and maximum distances between the projected planning structure and the delineated structure on the portal image. Setup time for each fraction was determined using 2 recorded time outs: one upon the patient entering the treatment room and another before radiation beam on.
Results:
Setup errors for the skin were significantly reduced with SGRT for all 4 treatment techniques (P < .001). On average, the mean and maximum errors for the skin decreased from 3.5 mm to 2.3 mm (P < .001) and from 7.6 mm to 5.6 mm (P < .001), respectively. Setup errors for the CW were not significantly different for tangent treatments, but significantly different for locoregional treatments. For all patients, the average mean and maximum errors for the CW were reduced from 3.1 mm to 3.0 mm (P = .21) and from 6.1 mm to 5.5 mm (P = .001), respectively. No significant change in setup errors for the heart was observed. Setup times with SGRT were slightly longer (P < .01), and the average setup time increased from 5.4 to 6.3 minutes.
Conclusions:
Using anatomic landmarks, we confirm that SGRT improved patient setup accuracy with a slight, but clinically nonsignificant increase in setup time.
More Related Videos
Related Concept Videos
Flail Chest-I
Flail chest is a severe and potentially life-threatening condition characterized by the fracture of three or more adjacent ribs in multiple places. It is most commonly caused by direct impacts and trauma, such as motor vehicle accidents or injuries from a steering wheel impact. It can also occur due to falls in elderly individuals with osteoporosis, or assaults involving sharp objects.
Pathophysiology
The pathophysiology of flail chest is complex, involving fractures of...
Flail Chest-II
Assessment:
1. Clinical Evaluation:
History:
Chest Physiotherapy
Purpose
CPT is primarily used for patients with excessive bronchial secretions who have difficulty clearing...
Biological Effects of Radiation
Improving Translational Accuracy
Uncertainty in Measurement: Accuracy and Precision