Related Experiment Video
Updated: Jan 27, 2026

Measurement of Tissue Oxygenation Using Near-Infrared Spectroscopy in Patients Undergoing Hemodialysis
Published on: October 2, 2020
Anaphylactic reaction to ethylene oxide in a hemodialysis patient
Bobak J Akhavan1, Ugochi A Osborn2, Reeba Mathew3
1Department of Internal Medicine, McGovern Medical School, The University of Texas Health Science Center at Houston, Houston, TX, USA.
Abstract:
We present the case of a patient who developed a severe systemic allergic reaction during initiation of hemodialysis. The reaction completely resolved by switching the dialysis filter sterilized by ethylene oxide to a steam sterilized filter. Ethylene oxide is used to sterilize heat sensitive medical devices, and although allergic reactions related to ethylene oxide have been reported before, awareness is lacking among providers in the inpatient setting, specifically in the intensive care unit setting.