Successful repositioning of apically tethered transcatheter mitral valve replacement in the off-bypass, beating heart
Dincer Aktuerk1, Paul Jansz1, Marty Shaw1
1Department of Cardiothoracic Surgery, St. Vincent's Hospital, Sydney, Australia.
Abstract:
A 73-year-old man with severe, symptomatic secondary mitral regurgitation (MR) underwent successful transcatheter mitral valve replacement using the Tendyne™ mitral valve system. The device was deployed from the left ventricular (LV) apex and secured in position by a tether attached to an epicardial pad. Three days postoperatively, the patient developed hemolytic anemia and a paravalvular leak (PVL) associated with indentation of the LV apex. Adjustment of the tether tension and placement of an epicardial disc under the pad resulted in reduction in the PVL, and resolution of the hemolytic anemia.
More Related Videos
05:31Transcatheter Pulmonary Valve Replacement from Autologous Pericardium with a Self-Expandable Nitinol Stent in an Adult Sheep Model
Published on: June 8, 2022
07:42An Image Guided Transapical Mitral Valve Leaflet Puncture Model of Controlled Volume Overload from Mitral Regurgitation in the Rat
Published on: May 19, 2020
Related Concept Videos
Mitral Valve Prolapse I: Introduction
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Mitral Valve Prolapse II: Assessment and Management
Mitral Valve Prolapse III: Nursing Management
Anatomy of the Heart
Beats