Radiofrequency transmit calibration: A multi-center evaluation of vendor-provided radiofrequency transmit mapping
Yannick Bliesener1, Xinran Zhong2, Yi Guo1
1Ming Hsieh Department of Electrical Engineering, Viterbi School of Engineering, University of Southern California, Los Angeles, CA, USA.
Purpose:
To determine the accuracy and test-retest repeatability of fast radiofrequency (RF) transmit measurement approaches used in Dynamic Contrast Enhanced Magnetic Resonance Imaging (DCE-MRI). Spatial variation in the transmitted RF field introduces bias and increased variance in quantitative DCE-MRI metrics including tracer kinetic parameter maps. If unaccounted for, these errors can dominate all other sources of bias and variance. The amount and pattern of variation depend on scanner-specific hardware and software.
Methods:
Human tissue mimicking torso and brain phantoms were constructed. RF transmit maps were measured and compared across eight different commercial scanners, from three major vendors, and three clinical sites. Vendor-recommended rapid methods for RF mapping were compared to a slower reference method. Imaging was repeated at all sites after 2 months. Ranges and magnitude of RF inhomogeneity were compared scanner-wise at two time points. Limits of Agreement of vendor-recommended methods and double-angle reference method were assessed.
Results:
At 3 T, B1 + inhomogeneity spans across 35% in the head and 120% in the torso. Fast vendor provided methods are within 30% agreement with the reference double angle method for both the head and the torso phantom.
Conclusions:
If unaccounted for, B1 + inhomogeneity can severely impact tracer-kinetic parameter estimation. Depending on the scanner, fast vendor provided B1 + mapping sequences allow unbiased and reproducible measurements of B1 + inhomogeneity to correct for this source of bias.
Related Concept Videos
Sexually Transmitted Infections
NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Lattice Centering and Coordination Number
Types of Unit Cells
Imagine taking a large number of identical...
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Center of Gravity