ELISA methods comparison for the detection of auto-antibodies against apolipoprotein A1
Miguel A Frias1, Julien Virzi1, Joana Batuca2
1Division of Laboratory Medicine, Diagnostic Department, Geneva University Hospitals, 4 rue Gabrielle-Perret-Gentil, 1205 Geneva, Switzerland; Department of Medical Specialties, Faculty of Medicine, University of Geneva, 1 rue Michel Servet, 1205 Geneva, Switzerland.
Background:
Autoantibodies against apolipoprotein A1 (anti-apoA1 IgG) have emerged as an independent biomarker for cardiovascular disease and mortality. Across studies, different ELISA methods have been used to measure the level of circulating anti-apoA1 IgG which could lead to substantial result differences between assays.
Objectives:
To make a comparative study of available anti-apoA1 IgG detection methods and to determine whether the choice of matrix sample (serum vs plasma) could influence the results.
Methods:
Blood samples were obtained from 160 healthy blood donors and collected on 4 different matrixes (serum, plasma-EDTA, -citrate, -lithium-heparinate). Anti-apoA1 IgG was measured using two homemade (Geneva's and Lisbon's) and one commercial ELISA kits. Passing-Bablok and Bland-Altman were used to compare the results. Anti-apoA1 IgG seropositivity cut-offs were defined according to the user's/manufacturer's criterion.
Results:
The current results showed substantial differences between those 3 assays. The dynamic ranges were significantly different, the commercial kit displaying the narrowest one. Passing-Bablok analysis demonstrated important proportional and constant biases between assays. The anti-apoA1 IgG seropositivity rate in Geneva, Lisbon and commercial assays varied between 24.5% and 1.9%. Matrix comparisons demonstrated that the matrix choice (plasma versus serum) influenced anti-apoA1 IgG results as well as the seropositivity rate in an assay-dependent manner. The coating antigen source was identified as important factor underlying results heterogeneity across assays.
Conclusions:
These results highlight the impact of the method and the cut-off used on anti-apoA1 IgG results and emphasize the need of standardizing existing assays. Given the important matrix influence, we suggest to use serum as matrix of choice.
Related Concept Videos
Renal Drug Clearance: Comparison Between Renal Excretion Methods
Renal clearance is often associated with the renal glomerular filtration rate (GFR), which represents the rate at which plasma is filtered through the glomeruli in the kidney. When drug reabsorption is minimal and there is no active secretion, renal clearance is closely related to the...
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a...
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Antibody Structure
Antibodies, also known as immunoglobulins (Ig), are essential players of the adaptive immune system. These antigen-binding proteins are produced by B cells and make up 20 percent of the total blood plasma by weight. In mammals, antibodies fall into five different classes, which each elicits a different biological response upon antigen binding.
The Y-Shaped Structure of Antibodies Consists of Four Polypeptide Chains
Antibodies consist of four polypeptide chains: two identical heavy...
Precipitation Titration: Endpoint Detection Methods
In the Volhard method, a standard excess of AgNO3 is first added to the...