Mwila Mulubwa1, Pierre Mugabo1
1School of Pharmacy, University of the Western Cape, Private Bag X17, Bellville 7535, Cape Town, South Africa.
Abstract:
A simple and sensitive ultra-performance liquid chromatography tandem mass spectrometry method has been developed and validated for the analysis of cycloserine in patients' plasma. Using methanol, cyloserine and propranolol (internal standard (IS)) was extracted from plasma by protein precipitation procedure. The chromatographic separation was successfully achieved on Phenomenex KinetexTM PFP C18 (2.1 mm × 100 mm, 2.6 μm) reversed-phase column. Acidified with 0.1% formic acid, water and acetonitrile were used as mobile phases for gradient elution. Cycloserine and IS were detected by Xevo® TQ MS triple quadrupole tandem mass spectrometer. The transition of protonated precursor to product ion were monitored at 103 → 75 m/z and 260.2 → 183 m/z for cycloserine and IS, respectively. The lower limit of quantification was 0.01 μg/mL. The method was linear over the concentration range 0.01-50 μg/mL with average coefficient of determination of 0.9994. The within-run and between-run precision and accuracy were in the range 3.7-19.3% (RSD) and 98.7-117.3%, respectively. Processed cycloserine sample was stable for 48 hours at 8°C and after three freeze-thaw cycles. The extraction efficiency ranged between 88.7 and 91.2%. The method was successfully applied in a pharmacokinetic study for the determination of cycloserine in plasma of patients with drug-resistant tuberculosis.
More Related Videos
08:56Quantitative Analysis of the Cellular Lipidome of Saccharomyces Cerevisiae Using Liquid Chromatography Coupled with Tandem Mass Spectrometry
Published on: March 8, 2020
09:09Semi-Quantitative Analysis of Peptidoglycan by Liquid Chromatography Mass Spectrometry and Bioinformatics
Published on: October 13, 2020
Related Concept Videos
Tandem Mass Spectrometry
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Instrumentation
Peptide Identification Using Tandem Mass Spectrometry
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...
High-Performance Liquid Chromatography: Elution Process
High-Performance Liquid Chromatography: Types of Detectors