Related Experiment Video
Updated: Jan 27, 2026

Measurement of Bioelectric Current with a Vibrating Probe
Published on: January 4, 2011
Piezospectroscopic measurement of high-frequency vibrations in a pulse-tube cryostat
Anne Louchet-Chauvet1, Rose Ahlefeldt2, Thierry Chanelière1
1Laboratoire Aimé Cotton, CNRS, Univ. Paris-Sud, ENS-Cachan, Université Paris-Saclay, 91405 Orsay, France.
Abstract:
Vibrations in cryocoolers are a recurrent concern to the end user. They appear in different parts of the acoustic spectrum depending on the refrigerator type, Gifford McMahon or pulse-tube, and with a variable coupling strength to the physical system under interest. Here, we use the piezospectroscopic effect in rare-earth doped crystals at a low temperature as a high resolution, contact-less probe for the vibrations. With this optical spectroscopic technique, we obtain and analyze the vibration spectrum up to 700 kHz of a 2 kW pulse-tube cooler. We attempt an absolute calibration based on known experimental parameters to make our method partially quantitative and to provide a possible comparison with other well-established techniques.