Related Experiment Video
Updated: Jan 27, 2026

Assessing the Accuracy of Fitness Smartwatch Data for Cardiovascular and Physical Activity Monitoring: A Validation Study in Digital Health
Published on: February 21, 2025
Assessment of agreement and interchangeability between the TEG5000 and TEG6S thromboelastography haemostasis
P Lloyd-Donald1, L Churilov2, F Zia1
1Department of Anaesthesia, Austin Hospital, 145 Studley Rd, Heidelberg, Victoria, 3084, Australia.
Background:
TEG6S® and TEG5000® (Haemonetics Corp, USA) are haemostasis analysers that measure viscoelasticity properties of whole blood. Both use different mechanisms to assess similar components of the coagulation process. The aim of this study was to assess agreement and interchangeability between the TEG6S and TEG5000 analysers.
Methods:
3.5 mL whole blood was collected from 25 adult patients in a tertiary intensive care unit (ICU). Analysis was performed using TEG6S and TEG5000 haemostatic platforms. Agreement between platforms was measured using Lin's concordance coefficient (Lin's CC), further validated using intraclass correlation coefficients and reduced major axis regression (RMAR).
Results:
Sixteen (64%) patients were male; mean (range) age: 59yo (23-86). TEG6S and TEG5000 systems were broadly interchangeable. The majority of TEG variables demonstrated almost perfect or substantial agreement and minimal proportional bias (maximum amplitude demonstrated a fixed bias). LY30%, however, demonstrated poor agreement and a proportional bias. Lin's CC coefficients (95% CI, RMAR slope, intercept) between TEG6S and TEG5000 variables were: R time: 0.78 (0.64-0.92, 0.76, 0.92); K time: 0.82 (0.69-0.94, 1.30, - 0.93); alpha angle: 0.79 (0.64-0.95, 1.04, - 1.43); maximum amplitude (MA): 0.90 (0.83-0.96, 0.99, - 5.0); LY30%: 0.34 (0.1-0.58, 0.43, 0.04).
Conclusions:
Adult patients with critical illness demonstrate almost perfect agreement in the R time and MA, substantial agreement in K time and alpha angle, but poor agreement in LY30%, as measured by the TEG6S and TEG5000 analysers. With the exception of LY30%, the TEG6S and TEG5000 platforms appear interchangeable. This has important implications for use in clinical practice and multi-site research programs.
Trial Registration:
ANZCRT number: 12617000062325 , registered 12/Jan17. Retrospectively registered.
Related Concept Videos
Protein Complexes with Interchangeable Parts
The SCF ubiquitin ligase is a protein complex of five individual proteins. This complex attaches ubiquitin to other target proteins to mark them for degradation. In order...
Protein Complexes with Interchangeable Parts
Reliability and Validity
Drug Products: Biologics, Biosimilars and Interchangeables
Data Validation
Key parameters for method validation include:
Data Validation
Nursing assessment guides are generally based on holistic models rather than medical...