Perturbation effect of parallel-plate ionization chambers on buildup dose measurements in transverse magnetic fields
Takanori Matsuoka1, Fujio Araki2, Takeshi Ohno2
1Graduate School of Health Sciences, Kumamoto University, 4-24-1 Kuhonji, Kumamoto, Japan.
Abstract:
The aim of this study is to investigate the perturbation effect of parallel-plate ionization chambers on the buildup dose measurement in transverse magnetic fields, using Monte Carlo (MC) simulation. The NACP-02 and ROOS parallel-plate chambers and a PTW31010 cylindrical chamber were modeled for buildup dose measurement in magnetic fields, using the EGSnrc/cavity code. The irradiation condition was set to a 10 × 10 cm2 field in a water phantom at a source-to-surface distance (SSD) of 100 cm, using 6-MV photon spectrum. Magnetic fields of 0 0.35, 1.0, 1.5, and 3.0 T were applied perpendicularly to the direction of the photon beam. The overall perturbation factor PQ,B for the ionization chambers in the magnetic fields was also calculated. The dose to water was enhanced with increasing the magnetic field strength at a depth of less than 1 cm. Over a depth of 1.5 cm, there was no significant difference in the depth doses with and without magnetic field in water. The maximum depth dose (%) for the NACP-02 and ROOS chambers at 1.5 T was higher up to 12% and 14% than the maximum depth dose at 0 T, respectively. The depth dose curves of a PTW31010 chamber have a similar tendency to those of water. The PQ,B values for each chamber were the largest at the phantom surface. The transverse magnetic field has a greater effect on the dose response of the NACP and ROOS chambers than that of the PTW31010 chamber in the buildup region.
More Related Videos
08:50Gene Expression Analysis of Endothelial Cells Exposed to Shear Stress Using Multiple Parallel-plate Flow Chambers
Published on: October 21, 2018
12:43Parallel-plate Flow Chamber and Continuous Flow Circuit to Evaluate Endothelial Progenitor Cells under Laminar Flow Shear Stress
Published on: January 17, 2012
Related Concept Videos
Electric Field of Parallel Conducting Plates
Consider a cross-section of a thin, infinite conducting plate having a positive charge. For such a large thin plate, as the thickness of the plate tends to zero, the positive charges lie on the plate's two large faces. Without an external electric field, the...
Steady, Laminar Flow Between Parallel Plates
Magnetic Fields
A magnetic field is defined by the force that a charged particle experiences...
Magnetic Force Between Two Parallel Currents
The force exerted by the magnetic field due to the first conductor over a finite length of the second conductor is given as the product of the current in the second conductor and the vector product of the length vector along the current element and the field due to the first conductor. According to the...
Magnetic Field of a Solenoid
Consider a solenoid with 100 turns wrapped around a cylinder of...
Magnetic Field Lines
Magnetic field lines follow several hard-and-fast rules: