Patient organ doses from megavoltage computed tomography delivery with a helical tomotherapy unit using a general
Hironori Nagata1,2, Satoru Sugimoto1, Hideyuki Hongo2
1Department of Radiation Oncology, Graduate School of Medicine, Juntendo University, Bunkyo-ku, Tokyo, Japan.
Abstract:
The purpose of this study was to quantify actual patient organ doses from megavoltage computed tomography (MVCT) using an MVCT beam model of a helical tomotherapy unit in a general treatment planning system (TPS). Dosimetric parameters (percentage depth dose, lateral beam profile, and longitudinal beam profile) of the MVCT beam were measured using Gafchromic EBT3 films (ISP Corporation, Wayne, NJ, USA) and used for beam modeling in a Pinnacle3 TPS (Philips, Amsterdam, Netherlands); this TPS is widely used with linear accelerators. The created beam model was adjusted and validated by assessing point doses in a cylindrical phantom in static and helical beam plans with fine, normal and coarse pitches. Maximum doses delivered to important organs from MVCT delivery for five clinical cases were calculated using the created beam model. The difference (average ± one standard deviation for all evaluation points) between calculated and measured doses was -0.69 ± 1.20% in the static beam plan. In the helical beam plan, the differences were 1.83 ± 2.65%, 1.35 ± 5.94% and -0.66 ± 8.48% for fine, normal and coarse pitches, respectively. The average maximum additional dose to important organs from MVCT in clinical cases was 0.82% of the prescribed dose. In conclusion, we investigated a method for quantifying patient organ dose from MVCT delivery on helical tomotherapy using an MVCT beam model in a general TPS. This technique enables estimation of the patient-specific organ dose from MVCT delivery, without the need for additional equipment.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Drug Dosing: Geriatric Patients
Drug Dosing: Obese Patients
Imaging Studies III: Computed Tomography
Nursing Process for Patient and Caregiver Teaching II: Planning and Implementation
Measurement: Standard Units