Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
Direct measurement of ion chamber correction factors, k Q and k B, in a 7 MV MRI-linac
Leon de Prez1, Simon Woodings, Jacco de Pooter
1VSL-Dutch Metrology Institute, Delft, The Netherlands. Department of Radiotherapy, University Medical Center Utrecht, Utrecht, The Netherlands. Author to whom correspondence may be addressed.
Abstract:
The output of MRI-integrated photon therapy (MRgXT) devices is measured in terms of absorbed dose to water, D w. Traditionally this is done with reference type ion chambers calibrated in a beam quality Q 0 without magnetic field. To correct the ion chamber response for the application in the magnetic field, a factor needs to be applied that corrects for both beam quality Q and the presence of the magnetic field B, k Q,B. This can be expressed as the product of k Q, without magnetic field, and ion chamber magnetic field correction, k B. k B depends on the magnetic field strength and its direction, the direction of the beam and the orientation and type of the ion chamber. In this study, for the first time, both k Q and k B were measured directly for six waterproof ion chambers (3 × PTW 30013 and 3 × IBA FC65-G) in a pre-clinical 7 MV MRI-linac at 0 T and at 1.5 T. Measurements were done with the only available primary standard built for this purpose, a water calorimeter. Resulting k Q factors for PTW and IBA chambers were 0.985(5) and 0.990(4), respectively. k B factors were measured with the chambers in antiparallel direction to the magnetic field (|| 180°), and perpendicular direction (⊥ -90°). k B|| and k B⊥ for the PTW chambers were 0.985(6) and 0.963(4), respectively and for IBA chambers 0.995(4) and 0.956(4). Agreement with the available literature values was shown, partly caused by the relatively large standard deviation (SD) in those values. The values in this study are currently the only available measured values for k Q and k B in an MRI-linac that are directly linked to the international traceability framework for the quantity absorbed dose to water, D w.
Related Concept Videos
Power Factor Correction
Voltammetry: Factors Affecting Measurements
Distance Corrections
Common Ion Effect
Factors Affecting Solubility
Electrolytes: van't Hoff Factor
The colligative properties of a solution depend only on the number, not on the identity, of solute species dissolved. The concentration terms in the equations for various colligative properties (freezing point depression, boiling point elevation, osmotic pressure) pertain to all solute species present in the solution. Nonelectrolytes dissolve physically without dissociation or any other accompanying process. Each molecule that dissolves yields one...