Applying a commercial atlas-based synthetic Computed Tomography algorithm to patients with hip prostheses for
Jonathan Wyatt1, Hazel McCallum1
1Northern Centre for Cancer Care, Newcastle Upon Tyne Hospitals NHS Foundation Trust, Newcastle, UK.
Background And Purpose:
Magnetic Resonance (MR)-only prostate radiotherapy has recently been clinically implemented using commercial synthetic Computed Tomography (sCT) algorithms. However patients with hip prostheses have been excluded from all MR-only research to date and assumed to require dedicated sCT algorithms. This study aimed to investigate the dosimetric accuracy of applying a commercial sCT algorithm, based on an atlas of patients without hip prostheses, to patients with prostheses.
Materials And Methods:
18 patients with unilateral hip prostheses received MR and CT scans in the radiotherapy position. sCTs were generated from the MR using a commercial algorithm. The clinical Volumetric Modulated Arc Therapy (VMAT) plan, consisting of partial arcs which avoided the prosthesis, was recalculated using the sCT and the dose distribution compared.
Results:
The mean isocentre dose difference was ΔD = (-0.4 ± 0.2)% (mean ± standard error of the mean (sem), range - 1.9%, 1.1%) and the mean differences in Planning Target Volume, bladder and rectum mean doses were ≤0.3%. The 3D global gamma pass rate with dose difference 1% and distance to agreement 1 mm within the body was ΓBody1/1 = (95.0 ± 0.5)% (sem) and within the 50% isodose volume, which excluded the prosthesis, was Γ50%1/1 = (98.5 ± 0.4)% (sem). The pass rate within the PTV was ΓPTV2/2 ≥ 99.7% for all patients, although for PTVs close (≤3.5 cm) to the prosthesis ΓPTV1/1 < 85% for three patients. The sCT did not accurately represent the prosthesis with a mean difference in radiological isocentre depth near the prosthesis of ΔdOutsideRad = (15.8 ± 2.6) mm (sem). However inside the treatment plan arc the difference was ΔdInsideRad = (-1.8 ± 0.5) mm (sem).
Conclusions:
Using a commercial prostate sCT algorithm for patients with unilateral hip prostheses is dosimetrically accurate (<0.5%) as long as the routine prosthesis-avoidance treatment planning approach is used and the PTV is >3.5 cm from the prosthesis. This suggests MR-only prostate radiotherapy can be extended to patients with hip prostheses without requiring a specific sCT algorithm.
More Related Videos
05:23Author Spotlight: Noninvasive Cerebral Blood Flow Determination in Human Functional Brain Region for Diagnosis of Neurological Disorders
Published on: May 31, 2024
06:08A Cognitive Fusion-guided Prostate Biopsy Using Multiparametric Magnetic Resonance Imaging and Transrectal Ultrasound
Published on: March 21, 2025
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Magnetic Resonance Imaging
Imaging Studies III: Computed Tomography
Atomic Nuclei: Magnetic Resonance
Nuclear Magnetic Resonance (NMR): Overview
NMR spectroscopy generates a spectrum where the characteristic absorption frequencies of the sample are...
Resonance