Olivier Zaro-Weber1,2,3, Hermann Fleischer1,2,3, Lucas Reiblich1,2,3
1Max Planck Institute for Neurological Research, Cologne, Germany.
Objective:
Accurate identification of the ischemic penumbra, the therapeutic target in acute clinical stroke, is of critical importance to identify patients who might benefit from reperfusion therapies beyond the established time windows. Therefore, we aimed to validate magnetic resonance imaging (MRI) mismatch-based penumbra detection against full quantitative positron emission tomography (15 O-PET), the gold standard for penumbra detection in acute ischemic stroke.
Methods:
Ten patients (group A) with acute and subacute ischemic stroke underwent perfusion-weighted (PW)/diffusion-weighted MRI and consecutive full quantitative 15 O-PET within 48 hours of stroke onset. Penumbra as defined by 15 O-PET cerebral blood flow (CBF), oxygen extraction fraction, and oxygen metabolism was used to validate a wide range of established PW measures (eg, time-to-maximum [Tmax]) to optimize penumbral tissue detection. Validation was carried out using a voxel-based receiver-operating-characteristic curve analysis. The same validation based on penumbra as defined by quantitative 15 O-PET CBF was performed for comparative reasons in 23 patients measured within 48 hours of stroke onset (group B).
Results:
The PW map Tmax (area-under-the-curve = 0.88) performed best in detecting penumbral tissue up to 48 hours after stroke onset. The optimal threshold to discriminate penumbra from oligemia was Tmax >5.6 seconds with a sensitivity and specificity of >80%.
Interpretation:
The performance of the best PW measure Tmax to detect the upper penumbral flow threshold in ischemic stroke is excellent. Tmax >5.6 seconds-based penumbra detection is reliable to guide treatment decisions up to 48 hours after stroke onset and might help to expand reperfusion treatment beyond the current time windows. ANN NEUROL 2019;85:875-886.
Insights
Magnetic resonance imaging (MRI) Tmax maps accurately identify the ischemic penumbra, the brain tissue at risk in stroke. This method reliably guides reperfusion therapy decisions up to 48 hours after stroke onset.
Area of Science:
- Neurology
- Radiology
- Medical Imaging
Background:
- Accurate identification of the ischemic penumbra is crucial for selecting acute stroke patients for reperfusion therapies.
- Current methods for penumbra detection face limitations, especially beyond established time windows.
Purpose of the Study:
- To validate magnetic resonance imaging (MRI) mismatch-based penumbra detection against quantitative positron emission tomography (PET).
- To optimize perfusion-weighted (PW) MRI measures for detecting the ischemic penumbra in acute ischemic stroke.
Main Methods:
- Ten patients with acute/subacute ischemic stroke underwent PW/diffusion-weighted MRI and quantitative 15O-PET within 48 hours.
- Validation involved comparing PW measures, including time-to-maximum (Tmax), against PET-defined penumbra using receiver-operating-characteristic analysis.
- A separate group of 23 patients was used for comparative validation.
Main Results:
- The PW Tmax map demonstrated the best performance (area-under-the-curve = 0.88) in detecting penumbral tissue up to 48 hours post-stroke.
- An optimal threshold of Tmax >5.6 seconds achieved >80% sensitivity and specificity for discriminating penumbra from oligemia.
Conclusions:
- The PW Tmax measure exhibits excellent performance for detecting the penumbral flow threshold in ischemic stroke.
- Tmax >5.6 seconds reliably identifies the penumbra, supporting treatment decisions up to 48 hours and potentially extending reperfusion therapy windows.
More Related Videos
10:20Automation of a Positron-emission Tomography PET Radiotracer Synthesis Protocol for Clinical Production
Published on: October 26, 2018
09:21Human Brown Adipose Tissue Depots Automatically Segmented by Positron Emission Tomography/Computed Tomography and Registered Magnetic Resonance Images
Published on: February 18, 2015
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Magnetic Resonance Imaging
Imaging Studies IV: Magnetic Resonance Imaging
Atomic Nuclei: Magnetic Resonance
Resonance