Stopping power data for high-energy photon beams.

P Andreo, A Brahme

    Physics in Medicine and Biology
    |August 1, 1986
    PubMed
    Summary

    This study precisely calculated Spencer-Attix stopping power ratios for radiation therapy dosimetry using Monte Carlo simulations. Results show high agreement with existing protocols and experimental data, improving accuracy in high-energy photon beam dosimetry.

    Related Concept Videos

    You might also read

    Related Articles

    Articles linked to this work by shared authors, journal, and citation graph.

    Sort by
    Same author

    Determination of consensus k <sub>Q</sub> values for megavoltage photon beams for the update of IAEA TRS-398.

    Physics in medicine and biology·2020
    Same author

    Correction factors for ionization chamber measurements with the 'Valencia' and 'large field Valencia' brachytherapy applicators.

    Physics in medicine and biology·2018
    Same author

    Poster - Thur Eve - 46: The upcoming international code of practice for small static photon field dosimetry.

    Medical physics·2017
    Same author

    Specification of Dose Delivery in Radiation Therapy. Recommendations by the Nordic Association of Clinical Physics (NACP).

    Acta oncologica (Stockholm, Sweden)·2017
    Same author

    Monte Carlo calculated and experimentally determined output correction factors for small field detectors in Leksell Gamma Knife Perfexion beams.

    Physics in medicine and biology·2015
    Same author

    Fluence correction factors for graphite calorimetry in a low-energy clinical proton beam: I. Analytical and Monte Carlo simulations.

    Physics in medicine and biology·2013

    Area of Science:

    • Medical Physics
    • Radiation Dosimetry
    • Computational Physics

    Background:

    • Accurate dosimetry is critical for effective radiation therapy.
    • Spencer-Attix stopping power ratios are essential for photon beam dosimetry.
    • Previous calculations had inconsistencies and lacked detailed analysis of influencing factors.

    Purpose of the Study:

    • To calculate Spencer-Attix stopping power ratios for high-energy photon beams.
    • To provide a more consistent calculation method.
    • To investigate the dependence of stopping power ratios on beam attenuation, electron contamination, depth, and field size.

    Main Methods:

    • Utilized the Monte Carlo method for precise calculations.
    • Calculated stopping power ratios as a function of beam attenuation properties.
    • Investigated the impact of electron contamination, depth, and field size.

    Main Results:

    • Stopping power ratios were calculated with improved consistency.
    • Results demonstrate agreement within approximately 1% with most dosimetry protocols.
    • Agreement with recent experimental data is better than 0.5%.

    Conclusions:

    • The Monte Carlo method provides accurate Spencer-Attix stopping power ratios.
    • The calculated ratios are reliable for high-energy photon beam dosimetry in radiation therapy.
    • The findings support enhanced accuracy in clinical radiation dosimetry protocols.