Manufacturing Abdominal Aorta Hydrogel Tissue-Mimicking Phantoms for Ultrasound Elastography Validation
Published on: September 19, 2018
Radial-Biased Tracking Method to Assess Tissue Displacement in Intravascular Ultrasound Sequences: A Phantom
Marco A V M Grinet1, Matheus Cardoso Moraes1
1Laboratory of Image and Signal Processing, Institute of Science and Technology, Universidade Federal de São Paulo, São José dos Campos, Brazil.
Objectives:
We created and evaluated a pixel-tracking method capable of accurately identify the displacement of tissue in intravascular ultrasound (IVUS) images.
Methods:
Our proposed pixel-tracking method assessed the horizontal and vertical displacement of tissue from a numerical phantom of IVUS sequences. The proposed tracking method is based on a block-matching framework, comparing 2 distinct frames within a selected region by normalized cross-correlation. Our method, specialized for IVUS applications, reduced the tracking area by implementing a limiting radius and a radial bias during the search.
Results:
The method was evaluated by using 54 numerical phantom image sequences from 9 distinct arterial models, resulting in different arteries with atherosclerotic plaques under a range of pressures. The ground truth reference coordinates of the tracked tissue were extracted from each numerical phantom sequence. Our results were compared to 8 other methods present in the literature. The mean absolute tracking errors ± SD for our method were 15.56 ± 19.46 and 13.04 ± 13.82 μm for the horizontal and vertical directions, respectively, between 2 subsequent frames, and 162.58 ± 305.93 and 102.22 ± 130.61 μm from lower to higher pressures in the range of 6 frames (n = 42,036).
Conclusions:
Our application-specific pixel-tracking method showed promising results and no statistically significant tracking error (P = .954), comparable to state-of-the-art methods present in the literature. Application-specific tracking methods have advantages over general methods by turning tissue-specific behavior into a directional bias in the tracking algorithm.
More Related Videos
07:21Magnetic Resonance Derived Myocardial Strain Assessment Using Feature Tracking
Published on: February 12, 2011
08:19Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences
Published on: May 17, 2018
Related Concept Videos
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps:
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation
Confirmation Biases
Hindsight Biases
Bias
In statistics, a sampling bias is created when a sample is collected from a population, and some members of the population are not as likely to be chosen as others (remember, each member...
Position and Displacement