Validation of a novel smartphone application-enabled, patient-operated skin barrier device
Erin E Grinich1, Anuja V Shah2, Eric L Simpson2
1School of Medicine, Oregon Health & Science University, Portland, Oregon.
Background:
Transepidermal water loss (TEWL) and surface capacitance measure skin barrier permeability and stratum corneum (SC) hydration, respectively, and are frequently utilized in atopic dermatitis clinical trials. Many barrier devices are costly and often used only in the academic setting. GPSkin is a low-cost, patient-operated device that measures both TEWL and SC hydration. This study aimed to test the reliability of GPSkin and assess its correlation with current industry standards.
Materials And Methods:
GPSkin was compared to the Biox AquaFlux (TEWL) and Courage-Khazaka Corneometer (SC hydration). Participants with healthy skin (n = 50) collected measurements with GPSkin in Trial 1 without any device education and in Trial 2 with additional instruction. In Trial 2, the investigator also performed measurements with GPSkin. Spearman's coefficients (rs ) were performed to assess device correlation. Intraclass correlation coefficients (ICC) were calculated to determine reliability.
Results:
Overall, GPSkin was moderately correlated with current industry device measurements for TEWL (Trial 1 rs :0.48; Trial 2 rs :0.40 participant, 0.34 investigator) and SC hydration (Trial 1 rs :0.63; Trial 2 rs :0.45). GPSkin demonstrated "good" test-retest reliability for both TEWL (ICC: 0.89) and SC hydration (ICC: 0.85) measurements when participants were provided with some device education. There was no difference in reliability between participants provided with device education and investigators.
Conclusion:
Based on these findings, we concluded that GPSkin provides reasonably precise and reliable measurements of SC hydration and TEWL as compared to current devices.
Related Concept Videos
Reliability and Validity
Data Validation
Nursing assessment guides are generally based on holistic models rather than medical...
Data Validation
Key parameters for method validation include:
Physiological Barriers
The blood endothelial barrier is the most porous of these. It allows all small ionized, un-ionized, and lipophilic molecules to pass through the endothelial lining into the interstitial space...
Operational Amplifiers
Vector Operations
A vector multiplied by a scalar value is called scalar multiplication. The result obtained is a new vector with a different magnitude. If the scalar is positive, the direction of the vector remains the same, but if it is negative, the direction of the vector is reversed. For example, the product of the mass and velocity yields the momentum.