Analysis of 18FDG PET/CT Imaging as a Tool for Studying Mycobacterium tuberculosis Infection and Treatment in Non-human Primates
Published on: September 5, 2017
NEMA NU-4 performance evaluation of a non-human primate animal PET
Chai Pei1, Feng Baotong1, Zhang Zhiming1
1Beijing Engineering Research Center of Radiographic Techniques and Equipment, Institute of High Energy Physics, Chinese Academy of Sciences, Beijing 100049, People's Republic of China.
Abstract:
The Eplus-260 primate PET is an animal PET imaging system developed by the Institute of High Energy Physics, Chinese Academy of Sciences, which is designed to image non-human primates, especially the brain of large non-human primates. The system consists of 48 block detectors arranged in two 24-sided rings with a ring diameter of 263 mm and an axial extent of 64 mm. Each block detector is composed of a 16 × 16 cerium-doped lutetium-yttrium orthosilicate crystal array with a pixel size of 1.9 × 1.9 × 10 mm3. This article presents a performance evaluation of the PET scanner according to the National Electrical Manufacturers Association NU-4 2008 standards. All measurements were made for an energy window of 360-660 keV and a coincidence timing window of 2 ns. In terms of the FWHM, the FBP reconstructed spatial resolution results in all three directions at the radial position of 5 mm were better than or approached to 2 mm, and remained below 3.0 mm within the central 5 cm diameter of the FOV. The peak absolute sensitivity of the scanner was measured 1.80%. For a monkey-sized phantom, the scatter fraction was 34.2% and the peak noise equivalent count rate (NECR) was 26.5 kcps at 64.3 kBq/cc. The overall imaging capabilities of the scanner were also assessed using in vivo imaging study of a rhesus macaque. The performance measurements demonstrate that the Eplus-260 primate PET scanner has the potential ability to obtain good quality and high-contrast images for non-human primates, especially the brain of large non-human primates and could be considered as one technologically advanced dedicated non-human primate PET scanner available today.
Related Concept Videos
Dipeptidyl Peptidase 4 Inhibitors
Tonicity in Animals
Tonicity in Animals
Animal Mitochondrial Genetics
Piaget's Stage 4 of Cognitive Development
Abstract Reasoning and Hypothetical-Deductive Thinking
Unlike the concrete operational...
Amides to Amines: LiAlH4 Reduction
Amide reduction requires two equivalents of the reducing agent, acting as a source of hydride ions. As shown in the figure, the reaction is initiated with a nucleophilic attack by the hydride ion at the carbonyl carbon to form a tetrahedral intermediate.