Nikhil S Choudhari1, Ashutosh Richhariya2, Vidya Wadke3
1V S T Glaucoma Center, Dr. Kallam Anji Reddy Campus.
Precis:
We checked 190 tonometers every month and repaired faulty ones. Calibration error (CE) frequency reduced from 23% to 0.6% at 1 year. Tonometers needing one or >1 CE repair differed in survival but not in age.
Purpose:
The purpose of this study was to report the outcomes of a comprehensive program to maintain calibration status of the Goldmann applanation tonometer.
Methods:
This prospective cohort study was carried out at 2 tertiary eye care referral centers. We included 190 slit-lamp mounted Goldmann applanation tonometers (Model AT 900 C/M). Health care providers (error checking and reporting) and clinical engineers (maintenance) participated. The team carried out CE check once a month, and repair of faulty tonometers, if any, within 24 hours. Failure of tonometer was defined as development of unacceptable CE beyond the third repair. The main outcome measures were the frequency of CE and survival function of the tonometer over 1 year.
Results:
The median age of the tonometers was 10.7 (range, 0.2 to 25.1) years. The total number of repairs was 86. The proportion (95% confidence interval) of faulty tonometers reduced from 23.1% (17.7, 29.6) in the first month to 0.6% (0.1, 3.3) at 1 year (P<0.01). The median age of the tonometer did not differ between those needing (n=63, 9.4 y) and not needing (n=127, 10.7 y; P=0.24) repair. All tonometers requiring 1 CE repair (n=49, 25.7%) survived until 1 year. The survival of tonometers requiring >1 CE repair (n=14, 7.3%) was 40% at 1 year.
Conclusions:
Our in-house program maintained 92.6% tonometers error free. Number of repairs rather than age determined the need for replacement/sending back the tonometer to the manufacturer. Our simple and easy to follow maintenance program has the potential for wide application.
Related Concept Videos
Predicting Reaction Outcomes
Outcomes of Glycolysis
Cellular respiration can occur aerobically (with oxygen) or anaerobically (without oxygen). In the presence of oxygen, cellular respiration starts with glycolysis and continues with pyruvate...
Maintenance of the ES Cell State
Guidelines for Writing Outcome
Patient outcomes reflect the patient's response to the goal rather than what the nurse aims to achieve. Terminology should be observable and measurable to avoid the reader's interpretation. The desired outcome should be realistic and achievable in the designated care timeframe. Expected outcomes should align with adjunctive therapies. The outcome should enhance care...
Self-Evaluation Maintenance Model
Cancer Stem Cells and Tumor Maintenance
Cancer stem cells are thought to originate from tissue-specific normal stem cells or progenitor cells. The normal stem cells usually reside in...