Insert conformity variation affects kinematics and wear performance of total knee replacements
Qida Zhang1, Zhenxian Chen2, Jing Zhang1
1State Key Laboratory for Manufacturing System Engineering, School of Mechanical Engineering, Xi'an Jiaotong University, No.28, Xianning West Road, Xi'an, 710049, China.
Background:
The insert conformity is a critical factor for successful total knee replacement which must be considered in design of the implant. However, the effects of conformity on knee kinematics and wear under physiological environment are often neglected in previous studies. The present study involved evaluating the biomechanics and wear performance with regard to different insert conformity in total knee replacement.
Methods:
Different tibial inserts with different sagittal and coronal conformity levels were created and analyzed using a previously developed wear prediction framework, coupling a patient-specific musculoskeletal multibody dynamics simulation, finite element and wear analysis. The contact mechanics, kinematics, and wear performance were compared during 10 million cycles of wear simulation.
Findings:
The findings revealed that the knee kinematics was affected by sagittal conformity design variables, which further influenced the wear of insert bearing surface. Additionally, kinematics and wear of artificial knee joint were much more sensitive to sagittal than coronal conformity of tibial insert. The lower sagittal conformity designs had lower wear rates, worn area and contact area. In turn, the wear of insert bearing surface also changed insert conformity, and further impacted on knee kinematics.
Interpretation:
The present study indicated that the sagittal conformity design of insert surface played a crucial role to improve contact mechanics and kinematics performance and minimize wear of total knee replacement. The optimization of insert conformity should be considered carefully in implant design and surgical procedures.
Related Concept Videos
Conformity
What is Variation?
The range, standard deviation, standard error, and variance are the different measures of variation.
Range: The range is the difference between its maximum and...
Variation
When independent and dependent variables are plotted on a scatter plot, the slope of a line is a value that describes the rate of change between the two...
Conservative Site-specific Recombination and Phase Variation
The recognition sites for Cre recombinase called LoxP...
Kinematic Equations - I
Knee Joint
A total of seven ligaments support the knee joint. The patellar ligament, which is also attached to the quadriceps femoris...