Does software optimization influence the radiologists' perception in low dose paediatric pelvic examinations?
H Precht1, D Waaler2, C B Outzen1
1Conrad Research Programme, Health Sciences Research Centre, UCL, Niels Bohrs Allé 1, 5230, Odense M, Denmark.
Purpose:
To investigate whether software optimisation can improve an observers' perception of image quality in low dose paediatric pelvic examinations.
Methods:
Twenty-five consecutive patients (3-7 years old) were referred for a pelvic digital radiography (DR) examination. They were prospectively enrolled in the study over a 3-month period. Images were taken at 80 kV and 2-4 mAs depending on pelvic thickness (9-15 cm). A small focal spot, 130 cm SID: 10 cm air gap and 1 mm Al and 0.2 mm Cu additional filtration were also utilised. Images were acquired on a Canon DR detector and optimised using five different combinations of the multi-frequency processing software (Canon DR system version NE, Version 7.1 with SPECTRA) to comply with the ALARA principle. Image quality was blindly evaluated using the subjective Visual Grading Analysis (VGA) by five experienced musculoskeletal radiologists, including the evaluation of six anatomical image quality criteria (scored from 1 to 5).
Results:
Consistently, the VGA results indicated that by using software optimised parameters, image quality was suitable for diagnosis in 48-71% of all images. Based on a VGC analysis all software optimised images did have significant better image quality then the one with just the clinical settings. Noise reduction was the software setting which influenced the image quality the most, area under the curve (AUC) of 0.8172 95%CI 0.7953-0.8375.
Conclusion:
Software optimisation improve the radiologists' perception of image quality and should thus be thoroughly considered within clinical practise. Noise reduction is the software parameter which has the greatest influence.
More Related Videos
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Ultrasonography
During an ultrasonography procedure, a handheld device called a...
Radiological Investigation I: X-ray and CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies II: Ultrasonography