Related Experiment Video
Updated: Jan 26, 2026

Adsorption Device Based on a Langatate Crystal Microbalance for High Temperature High Pressure Gas Adsorption in Zeolite H-ZSM-5
Published on: August 25, 2016
A reference high-pressure CO2 adsorption isotherm for ammonium ZSM-5 zeolite: results of an interlaboratory study
H G T Nguyen1, L Espinal1, R D van Zee1
11National Institute of Standards and Technology, Gaithersburg, MD USA.
Abstract:
This paper reports the results of an international interlaboratory study led by the National Institute of Standards and Technology (NIST) on the measurement of high-pressure surface excess carbon dioxide adsorption isotherms on NIST Reference Material RM 8852 (ammonium ZSM-5 zeolite), at 293.15 K (20 °C) from 1 kPa up to 4.5 MPa. Eleven laboratories participated in this exercise and, for the first time, high-pressure adsorption reference data are reported using a reference material. An empirical reference equation [n -surface excess uptake (mmol/g), P-equilibrium pressure (MPa), a = -6.22, b = 1.97, c = 4.73, and d = 3.87] along with the 95% uncertainty interval (U = 0.075 mmol/g) were determined for the reference isotherm using a Bayesian, Markov Chain Monte Carlo method. Together, this zeolitic reference material and the associated adsorption data provide a means for laboratories to test and validate high-pressure adsorption equipment and measurements. Recommendations are provided for measuring reliable high-pressure adsorption isotherms using this material, including activation procedures, data processing methods to determine surface excess uptake, and the appropriate equation of state to be used.
Related Concept Videos
Isothermal Processes
An ideal gas can also undergo isothermal expansion or compression.
For example, consider 1 mole of an ideal gas inside an isolated cylinder at initial volume V...
Analyte Adsorption and Distribution
Inertial Frames of Reference
Non-inertial Frames of Reference
5-Number Summary
In a box plot, the minimum and maximum data values represent the lower and upper whiskers in the graph, and the median is designated as the center of the box in the chart. The first quartile and third...
Chemical Shift: Internal References and Solvent Effects
The internal reference compound generally used in NMR spectroscopy is tetramethylsilane (TMS). TMS is preferred because it is chemically inert, soluble in NMR solvents, and easily removable. Also, the highly shielded methyl protons in TMS yield an intense...