Calibrating TrueBeam jaws by considering collimator walkout to improve the dose uniformity at abutting field
1Department of Radiation Oncology, Brody School of Medicine, East Carolina University, Greenville, NC, USA.
Abstract:
Jaw positions on a linear accelerator are calibrated to have accurate field size values over the range of jaw positions and to have excellent junctions when matching fields. It is sufficient to have field size accuracy on the order of a millimeter for most clinical applications but good junctions require submillimeter precision and accuracy in the jaw positioning. Presented is a method to measure collimator walkout with the MV imager and a mathematical model to determine an optimal origin for calibrating jaws on the TrueBeam accelerator. The calibration procedure uses the jaw position encoders which are sufficiently accurate and precise enough to achieve a homogeneous junction dose for abutting fields.
Related Concept Videos
Electric Field of a Non Uniformly Charged Sphere
Consider a non-uniformly charged sphere, for which the density of charge depends only on the distance from a point in space and not on the direction. Such a sphere has a spherically symmetrical charge distribution. Here, the electric...
Electric Potential Energy in a Uniform Electric Field
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Uniform Distribution
Two essential properties of this distribution are
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
P-N junction