Impact of Cine Frame Selection on Quantitative Coronary Angiography Results
Shigenori Ito1,2,3, Kanako Kinoshita1, Akiko Endo1
1Japan Cardiovascular Imaging Core Laboratory, Tokyo, Japan.
Abstract:
We evaluated intra- and interobserver variability of quantitative coronary angiography (QCA) due to cine frame selection for 9 coronary stenoses. The projection was selected in advance. Cine frames were selected by 2 blinded experts (blind frame QCA) followed by assignment by supervisor (pre-selected frame QCA). Each expert analyzed 18 frames twice with a 3-month interval. A total of 72 measurements by 2 experts were used for intra- and interobserver variability analysis in calibration factor (CF), minimal lumen diameter (MLD), percent diameter stenosis (%DS), interpolated reference diameter (Int R), and lesion length (LL). Accuracy, precision, and coefficient of variation (CV) were calculated based on 2 measurements. For interobserver variability, intraclass correlation coefficient (ICC) was evaluated. Regarding intraobserver variability, precision (CV) was 0.0026 (1.45), 0.220 (25.1), 0.282 (11.0), 7.626 (11.8), and 4.042 (28.7) for blind frame QCA and 0.0044 (2.46), 0.094 (11.2), 0.225 (8.6), 3.924 (5.9), and 1.941 (12.1) for pre-selected frame QCA and regarding interobserver variability, precision (CV) was 0.0037 (2.09), 0.271 (31.8), 0.307 (11.9), 10.10 (15.4), and 5.121 (39.5) for blind frame QCA and 0.0050 (2.82), 0.098 (11.4), 0.246 (9.5), 5.253 (8.0), and 2.857 (19.0) for pre-selected frame QCA in CF, MLD, Int R, %DS, and LL, respectively. Intraclass correlation coefficient of Int R was almost perfect in blind and pre-selected frame QCA. Intraclass correlation coefficient of MLD, %DS, and LL were substantial/lower by blind frame QCA and improved to almost perfect by pre-selected frame QCA. Blind cine film selection might affect intra- and interobserver variability, especially in MLD and LL. In the multiple linear regression analysis, blind frame QCA was selected as an explanatory factor of QCA variability in MLD, %DS, and LL. The error range due to frame selection must be taken into consideration in clinical use.
More Related Videos
06:39Ultrasound Based Assessment of Coronary Artery Flow and Coronary Flow Reserve Using the Pressure Overload Model in Mice
Published on: April 13, 2015
06:18Intravascular Ultrasound Image-Based Finite Element Modeling Approach for Quantifying In Vivo Mechanical Properties of Human Coronary Artery
Published on: December 6, 2024
Related Concept Videos
Framing Effects
Frames
Frames are versatile and widely used in various applications such as structural supports for beams and columns, automobile chassis construction, and in the construction...
Frames: Problem Solving I
Frames: Problem Solving II
Inertial Frames of Reference
Non-inertial Frames of Reference