Related Experiment Video
Updated: Jan 26, 2026

Investigating von Willebrand Factor Pathophysiology Using a Flow Chamber Model of von Willebrand Factor-platelet String Formation
Published on: August 14, 2017
Analytical Performance of a New Immunoturbidimetric Assay for von Willebrand Factor (VWF) Activity Testing
Background:
Von Willebrand disease requires laboratory confirmation with quantitative and qualitative measurements of von Willebrand factor (VWF). Qualitative VWF-activity (VWF-Ac) tests have poor inter- and intra-laboratory reproducibility with coefficients of variation (CVs) as high as 64%, often lacking accuracy at low VWF-Ac levels.
Methods:
This study evaluated the recently launched immunoturbidometric STAGO® STA-VWF:RCo® reagent for VWF-Ac. Accuracy was evaluated on 32 samples by comparing results using the Siemens® Innovance® reagent. An intra-run reproducibility study was performed on controls. Linearity and lower limit of detection was studied on external-quality-assurance (EQA) material with a known VWF-Ac level.
Results:
STA-VWF:RCo® reagent results were within clinical interpretation agreement with Siemens® Innovance®. The reproducibility study yielded % CVs of 8.41 for normal and 11.46 for abnormal controls and the assay was linear between 73 and 14.6% and remained linear to 2% with extrapolation.
Conclusions:
The STAGO® STA-VWF:RCo® reagent showed clinically meaningful accuracy and acceptable precision.
Related Concept Videos
Factors Affecting Activity Coefficient
The activity coefficient value for an ion is close to one when the solution has almost zero ionic strength, i.e., when the solution shows close to ideal behavior. As the ionic strength of the solution increases from 0 to 0.1 mol/L, a...
Development of Analytical Methods
Transcription Factors
Analyte Adsorption and Distribution
Jung's Analytical Theory
Graphical and Analytic Representation of Sinusoids
The first step is measuring the peak-to-peak value, which is twice the amplitude of the sinusoid. This provides information about the maximum voltage swing of the waveform.
Secondly, the period and angular frequency are determined. The period is the time taken for one complete cycle of the waveform, while...