Inappropriate activation of pacemaker magnet response mode by CPAP masks
Yosef Schwartz1, Jeremiah Wasserlauf2, Alan V Sahakian3
1Department of Internal Medicine, Northwestern University Feinberg School of Medicine, Chicago, Illinois.
Background:
While electromagnetic interference is widely recognized as a cause of noise detected by cardiac implanted electronic devices (CIEDs), we report on two cases of inappropriate activation of magnet response mode due to magnets in continuous positive airway pressure (CPAP) masks.
Methods:
Two patients with St. Jude Medical dual-chamber pacemakers (Assurity DR 2240; St. Jude Medical, St. Paul, MN, USA) were discovered on routine outpatient device interrogation to have magnet response events that correlated with nightly use of their CPAP masks. A magnet response event was reproduced by positioning one CPAP mask over the patient's pulse generator site. The magnetic flux density was measured using a Gauss meter.
Results:
It was found that their CPAP masks had magnetic clips that were used to secure the headgears. These appeared to be transiently positioned within sufficient proximity of the patients' CIEDs to activate magnet response events during sleep. The measured magnetic field density adjacent to a mask magnet was 136 Gauss.
Conclusion:
Clinicians managing CPAPs and CIEDs should be aware of magnetic interference with the potential for inappropriate activation of magnet response mode. CPAP masks that are secured with magnetic clips should be avoided in patients with CIEDs, since this can result in inappropriate asynchronous pacing in pacemakers or potential failure to deliver life-saving tachytherapies in patients with implanted cardioverter-defibrillators.
More Related Videos
09:01Magnetic-, Acoustic-, and Optical-Triple-Responsive Microbubbles for Magnetic Hyperthermia and Pothotothermal Combination Cancer Therapy
Published on: May 22, 2020
11:02Combining Transcranial Magnetic Stimulation and fMRI to Examine the Default Mode Network
Published on: December 28, 2010
Related Concept Videos
What is a Mode?
There can be more than one mode in a data set if multiple values have the same highest frequency. For instance, suppose that the Statistics exam scores of 20 students are: 50; 53; 59; 59; 63; 63; 72; 72; 72; 72; 72; 76; 78; 81; 83; 84; 84; 84; 90; 93. Here, the mode is 72, as it occurs most frequently, five times.
A data set with two modes is called bimodal. For example,...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Masking and Demasking Agents
There are many masking agents, such as cyanide, fluoride, triethanolamine, thiourea, and 2,3-bis(sulfanyl)propan-1-ol (formerly 2,3-dimercapto-1-propanol), with the masking agent chosen based on...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
MOSFET: Enhancement Mode
In their basic form, enhancement-mode MOSFETs are typically non-conductive when the gate-source voltage (Vgs) is zero. This default 'off' state means no...