Assessment of Vascular Function in Patients With Chronic Kidney Disease
Published on: June 16, 2014
Comparisons between different blood pressure measurement techniques in patients with chronic kidney disease
Shahrokh Ezzatzadegan Jahromi1, Ghasem Haghighi2, Jamshid Roozbeh1
1Shiraz Nephro-Urology Research Center, Shiraz University of Medical Sciences, Shiraz, Iran.
Background:
Automated office blood pressure (AOBP) machines measure blood pressure (BP) multiple times over a brief period. We aimed to compare the results of manual office blood pressure (MOBP) and AOBP methods with ambulatory BP monitoring (ABPM) in patients with chronic kidney disease (CKD).
Methods:
This study was performed on 64 patients with CKD (stages 3-4). A nurse manually measured the BP on both arms using a mercury sphygmomanometer, followed by AOBP of the arm with the higher BP and then ABPM. Mean BP readings were compared by paired t test and Bland-Altman graphs.
Results:
The mean ± standard deviation (SD) age of participants was 59.3 ± 13.6. The mean ± SD awake systolic BP obtained by ABPM was 140.2 ± 19.0 mmHg, which was lower than the MOBP and AOBP methods (156.6 ± 17.8 and 148.8 ± 18.6 mmHg, respectively; P < 0.001). The mean ± SD awake diastolic BP was 78.6 ± 13.2 mmHg by ABPM which was lower than the MOBP and AOBP methods (88.9 ± 13.2 and 84.1 ± 14.0 mmHg, respectively; P < 0.001). Using Bland-Altman graphs, MOBP systolic BP readings showed a bias of 16.4 mmHg, while AOBP measurements indicated a bias of 8.6 mmHg compared with ABPM.
Conclusion:
AOBP methods may be more reliable than MOBP methods for determining BP in patients with CKD. However, the significantly higher mean BPs recorded by AOBP method suggested that AOBPs may not be as accurate as ABPM in patients with CKD.
Related Concept Videos
Chronic Kidney Disease I: Introduction
Chronic Kidney Disease II: Clinical Manifestations
Measurement of Blood Pressure
Chronic Kidney Disease III: Interprofessional Care
Chronic Kidney Disease IV: Nursing Management
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...