Influenza B false-positive results by rapid molecular tests Alere™ i Influenza A&B 2 in France
Related Experiment Video
Updated: Jan 26, 2026

Rapid Molecular Detection and Differentiation of Influenza Viruses A and B
Published on: January 30, 2017
Marie Pierre Otto1, Anne-Lise Toyer2, Cécile Poggi2
1Service de microbiologie, Hôpital d'Instruction des Armées Sainte Anne, Toulon, France.
Abstract:
The Alere-i™ Influenza A&B (Abbott), a nicking endonuclease amplification reaction test, has recently been improved in order to deliver results in a few minutes. Our field observation highlights two problems with this new version: improper interpretation of a test as valid despite improper reagent hydration and falsely influenza B positive results. We advise users of the new system to check reagent hydration prior to reporting a result and to systematically confirm positive influenza B results.