Inaccuracies in Measuring Velocities and Timing of Flow and Tissue Motion Using High-End Ultrasound Systems
Andrew R Walker1, Tokuhisa Uejima2, Christian Prinz3
1Department of Biomedical Engineering, Västmanland Hospital, Västerås, Sweden.
Abstract:
To evaluate the accuracy of measurements of the velocities and timing of blood flow and of tissue motion, we tested four commercially available ultrasound systems. Pulsed wave, continuous wave and tissue Doppler recordings acquired across a range of angles of insonation were compared against a calibrated Doppler string phantom. Temporal delays were measured from the onset of the simulated electrocardiogram to the start of each Doppler signal. Across all modalities and angles, the mean errors of measurements of velocity using the four systems were 0.2%, 1.7%, 6.0% and 3.1%. The largest errors occurred using tissue Doppler at 5 cm/s (mean overestimation: 5.8%, range: +1.1%-12.5%). Mean delays in displaying the onset of flow were -3, 6, 11 and -16 ms. All differences between machines were statistically significant. The accuracy of high-end ultrasound systems for measuring velocities is better than in earlier reports, but the reporting of routine testing against standard phantoms should be mandatory.
Related Concept Videos
Relative Motion Analysis - Velocity
When an external force is exerted, it sets the crank into a rotational movement. This, in turn, instigates the motion of the connecting rod, leading to what is referred to as a general plane motion. This process involves two key points - point A on the connecting rod...
Velocity and Acceleration in Steady and Unsteady Flow
The acceleration can be generalized to any point in the flow, and expressed as components along three perpendicular directions, representing changes in velocity over...
Linear time-invariant Systems
The input-output behavior of an LTI system can be fully defined by its response to an impulsive excitation at its input. Once this impulse response is known, the system's reaction to any other input can be...
BIBO stability of continuous and discrete -time systems
To determine the BIBO stability, the convolution integral is utilized when a bounded continuous-time input is applied to a Linear Time-Invariant (LTI) system....
Average Velocity
Instantaneous Velocity - II