The influence of CT and dual-energy X-ray absorptiometry (DXA) bone density on quantitative [18F] sodium fluoride PET
Inderbir S Jassel1, Musib Siddique1, Michelle L Frost2
1School of Biomedical Engineering & Imaging Sciences, King's College London, St Thomas' Hospital, London, UK.
Background:
[18F] sodium fluoride PET/CT provides quantitative measures of bone metabolic activity expressed by the parameters standardised uptake value (SUV) and bone plasma clearance (K i) that correlate with measurements of bone formation rate obtained by bone biopsy with double tetracycline labelling. Both SUV and K i relate to the tracer uptake in each millilitre of tissue. In general, the bone region of interest (ROI) includes both mineralised bone {generally with a high concentration of [18F]NaF} and bone marrow (with a much lower concentration), suggesting that correcting SUV and K i for volumetric bone mineral density (vBMD) and measuring them with respect to the tracer uptake in each gram of bone mineral might improve the correlation with the findings of bone biopsy. As a first test of this hypothesis, we looked for positive correlations between SUV and K i values with CT and DXA bone mineral density (BMD) parameters measured in the same ROI.
Methods:
A retrospective reanalysis was performed of 63 lumbar spine [18F]NaF PET/CT scans acquired in four earlier studies. The quantitative PET parameters SUV and K i were measured in L1-L4 and Hounsfield units (HU) measured on the CT scans in the same ROI. Spine BMD data was also obtained from DXA scans in the form of areal BMD and used to derive the bone mineral apparent density (BMAD, an estimate of vBMD). Scatter plots were drawn of SUV and K i against HU, BMAD and areal BMD and the Spearman rank correlation coefficients derived for each plot.
Results:
All correlations were positive and statistically significant. Correlations were highest for HU (SUV: RS =0.513, P<0.0001; K i: RS =0.429, P=0.0005) and lowest for areal BMD (SUV: RS =0.353, P=0.005; K i: RS =0.274, P=0.03).
Conclusions:
The results demonstrate significant positive correlations between SUV and K i and vBMD measurements in the form of HU from CT or BMAD and areal BMD from DXA. These findings justify further exploration of the relationship between SUV and K i [18F]NaF PET/CT measurements and CT or DXA measurements of vBMD to examine whether normalization for bone density might improve their correlation with bone metabolic activity as measured by bone biopsy.
More Related Videos
15:10A Dual Tracer PET-MRI Protocol for the Quantitative Measure of Regional Brain Energy Substrates Uptake in the Rat
Published on: December 28, 2013
06:31Quantitative [18F]-Naf-PET-MRI Analysis for the Evaluation of Dynamic Bone Turnover in a Patient with Facetogenic Low Back Pain
Published on: August 8, 2019
Related Concept Videos
Radiological Investigation I: X-ray and CT
Strain-Energy Density
In the elastic region of a material, the relationship between the stress and the strain is linear and follows Hooke's Law. The strain energy density in this region...
Trends in Lattice Energy: Ion Size and Charge
What is Energy?
X-ray Crystallography
Diffraction
Diffraction is the change in the direction of travel experienced by an electromagnetic wave when it encounters a physical barrier whose dimensions are comparable to those of the wavelength of the light. X-rays are electromagnetic radiation with wavelengths about as long as the distance between neighboring...
Energy Basics