Comparison of titer results obtained using immediate spin one-dilution techniques to a reference method
Mark H Yazer1,2, Nancy M Dunbar3, Carolina Bonet Bub4
1Department of Pathology, University of Pittsburgh, Pittsburgh, Pennsylvania.
This study evaluated immediate spin (IS) one-dilution titers for screening plasma products. Manual and automated IS buffered gel card methods showed high accuracy, making them suitable for transfusion services.
Area of Science:
- Transfusion Medicine
- Immunology
- Clinical Laboratory Science
Background:
- Transfusion services commonly use one-dilution titers to assess incompatible plasma-containing products.
- Evaluating the accuracy of immediate spin (IS) one-dilution titer methods is crucial for blood product safety.
Purpose of the Study:
- To compare the performance of three different immediate spin (IS) one-dilution titer methods against a reference standard method.
- To determine the suitability of these methods for screening plasma-containing products in transfusion services.
Main Methods:
- Three immediate spin (IS) one-dilution titer methods were evaluated: manual IS tube, manual and automated IS buffered gel card, and automated microplate.
- These methods were compared against a reference standard method involving saline tube testing with a 1-hour room temperature incubation.
- No time or temperature incubations were performed for the routine IS methods, and antihuman globulin reagent was not used in any method.
Main Results:
- The manual IS tube method demonstrated a sensitivity of 0.88 and specificity of 1.00.
- The manual and automated IS buffered gel card method showed a sensitivity of 0.95 and specificity of 0.87.
- The automated microplate IS method had a sensitivity of 0.76 and specificity of 0.96.
- All tested methods achieved an overall accuracy rate of 90% or greater.
Conclusions:
- Both manual and automated one-dilution titer methods are effective for screening plasma-containing units.
- Further evaluation of the automated microplate method is recommended for optimal use in transfusion services.
Related Concept Videos
Solution Concentration and Dilution
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Spin–Spin Coupling Constant: Overview
Qualitatively, any spin plus-half nucleus polarizes the spins of its electrons to the minus-half state. Consequently, the paired electron in the hydrogen–carbon bond must...
NMR Spectroscopy: Spin–Spin Coupling
Spin–Spin Coupling: One-Bond Coupling
Renal Drug Clearance: Comparison Between Renal Excretion Methods
Renal clearance is often associated with the renal glomerular filtration rate (GFR), which represents the rate at which plasma is filtered through the glomeruli in the kidney. When drug reabsorption is minimal and there is no active secretion, renal clearance is closely related to the...