Effect of scan pattern on complete-arch scans with 4 digital scanners
Related Experiment Video
Updated: Jan 26, 2026

Measuring the Complete-arch Distortion of an Optical Dental Impression
Published on: May 30, 2019
Jason Latham1, Mark Ludlow2, Anthony Mennito2
1Orthodontic Resident, Department of Orthodontics, College of Dental Medicine, Medical University of South Carolina, Charleston, SC.
Statement Of Problem:
Complete-arch digital scans are becoming popular as digital dentistry is adopted for expanded clinical situations such as complete-arch prostheses, removable prostheses, extensive implant-supported treatment, and orthodontic aligners. Whether the scan pattern technique affects the trueness and precision of complete-arch scans and whether differences in accuracy exist among different scanners remain unclear. Furthermore, each manufacturer recommends a different scan pattern, but evidence of the superiority of the manufacturer's recommended pattern is lacking.
Purpose:
The purpose of this in vitro study was to determine whether the scan pattern affects the trueness, precision, and speed of complete-arch digital scans performed by using 4 different digital scanning systems.
Material And Methods:
A custom model used as the reference standard was fabricated with teeth having the same refractive index as dentin and enamel to simulate the natural dentition. The scan of the custom typodont was obtained by using an ATOS III Triple Scan 3D optical scanner. This study evaluated the CEREC Omnicam, Planmeca Emerald, Align iTero Element, and 3Shape TRIOS 3. Experimental scans were obtained from each of the 4 different digital scanning systems by using 4 unique scan patterns by experienced clinicians. Four experimental scans were acquired from each of the scanners by using 4 distinct scan patterns for a total of 16 scans for each scanner. Scan patterns 1 to 4 were based on the operator manuals for each different scanner. The scan time was recorded for each scan. All experimental scans were converted to standard tessellation language (STL) format, and a comprehensive metrology program, Geomagic Control X, was used to compare the reference standard scan with the experimental scans.
Results:
For trueness, the scanner (P<.001), scan pattern (P=.001), and their interaction (P<.001) were found to be significant. Overall, scan pattern 2 showed the highest average trueness and precision. Likewise, for overall scan pattern precision, the scanner, scan pattern, and their interaction were found to be significant (P<.001).
Conclusions:
Scan pattern affected trueness and precision for some scanners, but not for others. Differences exist in the complete-arch scan speed, trueness, and precision of individual scanners. Scan pattern can play an important role in the success of digital scanning.
More Related Videos
Related Concept Videos
Leaky Scanning
Scanning Electron Microscopy
Fundamental Principles
Accelerated...
Voltammetric Techniques: Linear-Scan (E vs Time)
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
The Arch of Aorta
Encircling the heart, the coronary arteries form a ring-like structure before...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...