High impedance alert with safety switching: An unreported hazard of hybrid pacing systems
Ankur A Karnik1, Robert H Helm1, Matthew D Gaskill1
1Evans Department of Medicine, Cardiovascular Medicine Section, Arrhythmia Service, Boston University School of Medicine, Boston, Massachusetts.
Abstract:
Several Boston Scientific pacemaker models have a known issue with intermittent oversensing of the minute ventilation sensor when paired with non-Boston Scientific leads. Several of our patients with these hybrid systems have had transient out of range impedances and oversensing after safety switching which we suspected may be related. A retrospective analysis of 395 patients who had pacemakers implanted between 2015-2017 found that transient out of range impedances with safety switching was present in 9% of Boston Scientific pacemakers paired with Abbott or Medtronic leads compared with 0% in other device-lead combinations (P = 0.0089). We postulate that the root cause of the minute ventilation oversensing and transient high impedance issue is the same, a header-lead interaction from low-level incompatibility. Recognizing this issue is critical to prevent unnecessary lead revisions or extractions as it can be prevented with a simple reprogramming of lead pace/sense configuration.
Related Concept Videos
Hazard Ratio
For example, in a clinical trial...
Switching of BJT
Cut-off Mode ("Off" State): In this state, both the emitter-base and collector-base junctions are...
Hazard Rate
Survey Safety
Hybrid Zones
Impedances and Admittance
Impedance is a measure of resistance to sinusoidal current flow in an AC circuit. Unlike their behavior in DC circuits, where inductors appear as short circuits and capacitors as open circuits, the behavior of these components in AC...