Biomechanical Analysis of Cortical Versus Pedicle Screw Fixation Stability in TLIF, PLIF, and XLIF Applications
Edward K Nomoto1, Guy R Fogel2, Alexandre Rasouli1
1Cedars-Sinai Spine Center, Los Angeles, CA, USA.
Study Design:
Cadaveric biomechanical study.
Objectives:
Medial-to-lateral trajectory cortical screws are of clinical interest due to the ability to place them through a less disruptive, medialized exposure compared with conventional pedicle screws. In this study, cortical and pedicle screw trajectory stability was investigated in single-level transforaminal lumbar interbody fusion (TLIF), posterior lumbar interbody fusion (PLIF), and extreme lateral interbody fusion (XLIF) constructs.
Methods:
Eight lumbar spinal units were used for each interbody/screw trajectory combination. The following constructs were tested: TLIF + unilateral facetectomy (UF) + bilateral pedicle screws (BPS), TLIF + UF + bilateral cortical screws (BCS), PLIF + medial facetectomy (MF) + BPS, PLIF + bilateral facetectomy (BF) + BPS, PLIF + MF + BCS, PLIF + BF + BCS, XLIF + BPS, XLIF + BCS, and XLIF + bilateral laminotomy + BCS. Range of motion (ROM) in flexion-extension, lateral bending, and axial rotation was assessed using pure moments.
Results:
All instrumented constructs were significantly more rigid than intact (P < .05) in all test directions except TLIF + UF + BCS, PLIF + MF + BCS, and PLIF + BF + BCS in axial rotation. In general, XLIF and PLIF + MF constructs were more rigid (lowest ROM) than TLIF + UF and PLIF + BF constructs. In the presence of substantial iatrogenic destabilization (TLIF + UF and PLIF + BF), cortical screw constructs tended to be less rigid (higher ROM) than the same pedicle screw constructs in lateral bending and axial rotation; however, no statistically significant differences were found when comparing pedicle and cortical fixation for the same interbody procedures.
Conclusions:
Both cortical and pedicle trajectory screw fixation provided stability to the 1-level interbody constructs. Constructs with the least iatrogenic destabilization were most rigid. The more destabilized constructs showed less lateral bending and axial rotation rigidity with cortical screws compared with pedicle screws. Further investigation is warranted to understand the clinical implications of differences between constructs.
Related Concept Videos
Self-Locking Screw
A square-threaded screw jack carrying a load is considered self-locking if the screw retains its position even after the moment applied to it is removed.
Screw: Problem Solving
To evaluate the...
Frictional Forces on Screws
A jack with a square-threaded screw is a mechanical device used to lift heavy loads by applying a force at its handle. When the force is applied, the screw turns, raising the load. The screw can...
Nuclear Stability
To hold positively charged protons together...
RNA Stability
Fixation and Sectioning
The simplest type of preparation is the wet mount, in which the specimen is placed in a drop of liquid on the slide. A liquid specimen can be directly deposited on the slide using a dropper. Solid specimens, such as skin scraping, can be placed on the slide before adding a drop of liquid to prepare the wet mount. Sometimes the liquid is simply water, but stains are often added...