Repeatability and Agreement of a Swept-Source Optical Coherence Tomography-Based Biometer IOLMaster 700 Versus a
Tommy C Y Chan1, Kelvin H Wan, Fang Yao Tang
1Department of Ophthalmology and Visual Sciences (T.C.Y.C., K.H.W., F.Y.T., Y.M.W., and C.C.), The Chinese University of Hong Kong, Kowloon, Hong Kong; Hong Kong Eye Hospital (T.C.Y.C.), Kowloon, Hong Kong; and Department of Mathematics and Statistics (M.Y.), Hang Seng Management College, New Territories, Hong Kong.
Purpose:
To compare the repeatability and agreement between a swept-source biometer and a Scheimpflug biometer in cataract patients.
Methods:
Three consecutive measurements were obtained using a swept-source biometer (IOLMaster 700) and a Scheimpflug biometer (AL-Scan) in 52 eyes of 52 patients. Keratometry, central corneal thickness (CCT), anterior chamber depth (ACD), axial length, and white-to-white (WTW) distance were recorded. Astigmatism values were transformed into vector components of J0 and J45. Intraoperator repeatability was analyzed using intraclass correlation coefficients (ICCs) and reproducibility coefficients (RCs). Agreement of measurements between the two devices was evaluated using the Bland-Altman method.
Results:
The IOLMaster 700 showed higher ICCs and lower RCs for the mean keratometry (Km) (P≤0.018), CCT (P≤0.027), and ACD (P≤0.001) measurements, whereas the AL-Scan showed higher ICC and lower RC for the J45 vector component of astigmatism at the 2.4-mm zone (P≤0.034). Both the devices had excellent repeatability (ICC=0.999) in axial length measurement. Systematic differences were found in Km, CCT, ACD, and WTW (P≤0.018) between the devices. The mean difference for Km was -0.196 and -0.144 D measured at the 2.4-mm zone and 3.3-mm zone, respectively. The corresponding mean difference for CCT, ACD, and WTW distance was 14.92 μm, -0.017 mm, and 0.283 mm, respectively. These differences led to a statistically significant but clinically insignificant difference in the prediction of intraocular lens power.
Conclusions:
This study showed significant differences in anterior segment measurement repeatability and agreement between a swept-source biometer and a Scheimpflug biometer in eyes with cataract.
More Related Videos
12:22Multimodal Volumetric Retinal Imaging by Oblique Scanning Laser Ophthalmoscopy oSLO and Optical Coherence Tomography OCT
Published on: August 4, 2018
06:19In Vivo Imaging of Cx3cr1gfp/gfp Reporter Mice with Spectral-domain Optical Coherence Tomography and Scanning Laser Ophthalmoscopy
Published on: November 11, 2017
Related Concept Videos
Base Excision Repair
The first step of...
DNA Base Pairing
Lewis Acids and Bases
A coordinate covalent bond (or dative bond) occurs when one of the atoms in the bond provides both bonding electrons. For example, a coordinate covalent bond occurs when a water molecule combines with a hydrogen ion to form a hydronium ion. A coordinate covalent bond also results when...
Weak Base Solutions
Ions as Acids and Bases
Salts are ionic compounds composed of cations and anions, either of which may be capable of undergoing an acid or base ionization reaction with water. Aqueous salt solutions, therefore, may be acidic, basic, or neutral, depending on the relative acid-base strengths of the salt’s constituent ions. For example, dissolving the ammonium chloride in water results in its dissociation, as described by the equation:
Acid–Base Equilibria: Activity-Based Definition of pH
In solutions of very low ionic strength—for example, pure water—the...