Contamination Risk During Fecal Management Device Removal: An In vitro, Simulated Clinical Use Study
Daniel G Metcalf1, Lawrence GTsai2, Corrine Williams3
1ConvaTec, Deeside, UK.
Abstract:
Fecal management devices (FMDs) are used to drain and contain fecal matter in incontinent, often acutely or critically ill patients to protect their skin as well as the environment from contamination. However, there is potential for contamination and resultant infection at various stages of FMD use.
Purpose:
This in vitro study was conducted to compare device removal factors and subsequent splash of simulated fecal matter of 3 different designs of FMDs using a simulated rectum.
Methods:
A Universal Test Machine was used to automatically measure removal forces (in newtons [N]) and tube extensions as the FMDs were pulled from the simulated rectum by the machine. Splash distance and quantity were measured using a splash-capture cylinder and image analysis software. Each device was tested 3 times. Two-sample t tests were conducted to examine statistical differences in removal forces, removal extensions, and splash areas.
Results:
The forces required to remove the FMDs from the simulated rectum were significantly lower for the device with a collapsible, donut-shaped retention balloon compared with the devices with a green, foldable, trumpet-shaped retention cuff and a foldable, spherical-shaped retention balloon (12.0 ± 0.3 N vs. 32.6 ± 4.3 N and 34.8 ± 3.1 N, respectively; P <.05). The extensions of the catheter tubing were significantly lower for the device with a collapsible, donut-shaped retention balloon compared with the devices with a green, foldable, trumpet-shaped retention cuff and a foldable, spherical-shaped retention balloon (32.0 ± 7.5 mm vs. 81.3 ± 9.1 mm and 105.2 ± 10.6 mm, respectively; P <.05). Simulated fecal matter was splashed over mean areas of 25.5 ± 16.1 cm2 and 27.3 ± 13.5 cm2 for the devices with a green, foldable, trumpet-shaped retention cuff and a foldable, spherical-shaped retention balloon, respectively; no splash was observed for the device with a collapsible, donut-shaped retention balloon.
Conclusion:
In vitro observations suggest contamination and potential infection risk during FMD removal from the patient are influenced by FMD design. Future in vitro and clinical studies assessing the infectious nature of effluent and methods for containment are warranted.
Related Concept Videos
Esophageal Perforation-II: Clinical Manifestations and Management
Clinical Manifestations:
Barrett Esophagus-II: Clinical Manifestations and Management
To diagnose Barrett's esophagus, healthcare providers often recommend an endoscopy for those showing symptoms of acid reflux. The procedure...
Esophageal Varices-II: Clinical Features and Management
In the initial assessment, a thorough review of the patient's medical history is vital to identify risk factors such as liver disease, alcohol...
Gastritis III: Clinical Manifestations and Management
Clinical manifestations of acute gastritis
The patient with acute gastritis may have a rapid onset of symptoms, such as epigastric pain or discomfort, dyspepsia, anorexia, hiccups, or nausea and vomiting, which can last from a few hours to a few days. Erosive or hemorrhagic gastritis may cause bleeding, which may manifest as blood in vomit or as...
Esophageal Strictures-II: Clinical Features and Management
Healthcare providers should gather a comprehensive medical history and conduct a physical examination for diagnosis. If esophageal stricture is...
Acute Pancreatitis II: Clinical Manifestations and Management