Successful defibrillation verification in subcutaneous implantable cardioverter-defibrillator recipients by
Mauro Biffi1, Matteo Ziacchi1, Andrea Angeletti1,2
1PoloCardio-Toraco-Vascolare, Azienda Ospedaliero-Universitaria di Bologna, Bologna, Italy.
Background:
The subcutaneous implantable cardioverter-defibrillator (S-ICD) is an effective alternative to the transvenous one. Defibrillation efficacy depends on maximum device output and on the optimal device location at device implantation.
Hypothesis:
We sought to investigate the defibrillation safety margin in real life clinical practice.
Methods:
We sought to understand what is the efficacy of induced ventricular fibrillation (VF) termination at S-ICD implantation using lower energies than the recommended 65 J.
Results:
Sixty-four consecutive S-ICD recipients underwent VF termination attempts at implantation with energies ranging from 20 to 50 J. Overall, VF termination occurred in 84% of patients with ≤40 J, in 88% with 45 J, and in 100% with 60 J. Intermuscular S-ICD placement was associated with 94% VF termination at ≤40 J. An ejection fraction <35% was associated to higher energy requirement for defibrillation; however, an intermuscular S-ICD placement conferred 90% defibrillation efficacy at 31 ± 5 J in this patients subset.
Conclusions:
This is a hypothesis-generating observation that prompts a methodologically correct investigation to prove that a 60 J output S-ICD can provide an adequate safety margin to terminate VF in clinical practice. This would enable superior device longevity and/or device downsizing for pediatric/small size patients.
More Related Videos
03:44A Unique Mouse Model for Quantitative Assessment of Biofilm Formation on Surgical Implants in Subcutaneous Abscess
Published on: June 6, 2025
08:38Renal Capsule Xenografting and Subcutaneous Pellet Implantation for the Evaluation of Prostate Carcinogenesis and Benign Prostatic Hyperplasia
Published on: August 28, 2013
Related Concept Videos
Self-Evaluation: Self-Enhancement and Self-Verification
Ecological Succession
Strategies of Self-Presentation II: Self-Verification
Ionization Energy
Shock Waves
When the source's speed approaches the speed of sound, constructive interference between successive wavefronts emitted by the source occurs immediately behind it. Initially, scientists believed that this constructive interference would result in such high...
What is Energy?